On the Resolution Capabilities and Limits of Fluorescence Lifetime Correlation Spectroscopy (FLCS) Measurements

机译:荧光寿命相关光谱(FLCS)测量的分辨能力和极限

获取原文
获取原文并翻译 | 示例
       

摘要

Quantitative tests were performed in order to explore the practical limits of FLCS. We demonstrate that: a) FLCS yields precise and correct concentration values from as low as picomolar to micromolar concentrations; b) it is possible to separate four signal components in a single detector setup; c) diffusion times differing only 25% from each other can be resolved by separating a two component mixture based on the different fluorescence lifetimes of both components; d) most of the inherent technical limitations of conventional FCS are easily overcome by FLCS employing a single detector channel confocal detection scheme.
机译:为了探索FLCS的实际限制,进行了定量测试。我们证明:a)FLCS产生从皮摩尔到微摩尔浓度的精确而正确的浓度值; b)可以在单个检测器设置中分离四个信号分量; c)可以通过基于两种组分的不同荧光寿命分离两种组分的混合物来解决彼此仅相差25%的扩散时间; d)通过采用单个检测器通道共焦检测方案的FLCS可以轻松克服常规FCS的大多数固有技术局限性。