首页> 外文期刊>Journal of Endodontics: Official Journal of American Association of Endodontists >Cyclic Fatigue Resistance of Reciproc, WaveOne, and WaveOne Gold Nickel-Titanium Instruments
【24h】

Cyclic Fatigue Resistance of Reciproc, WaveOne, and WaveOne Gold Nickel-Titanium Instruments

机译:往复式,WaveOne和WaveOne金镍钛仪器的循环疲劳强度

获取原文
获取原文并翻译 | 示例
       

摘要

Introduction: The purpose of this study was to compare the cyclic fatigue resistance of Reciproc R25 (VDW, Munich, Germany), WaveOne Primary (Dentsply Maillefer, Ballaigues, Switzerland), and WaveOne Gold Primary files (Dentsply Maillefer). Methods: Twenty Reciproc R25, 20 WaveOne Primary, and 20 WaveOne-Gold Primary instruments were-included in this study. The-cyclic fatigue tests were performed using a cyclic fatigue testing device, which has an artificial stainless steel canal with a 60 angle of curvature and a 5-mm radius of curvature. The files were randomly divided into 3 groups (group 1: Reciproc R25; group 2: WaveOne Primary; and group 3: WaveOne Gold Primary). All the instruments were rotated until fracture occurred, and the time to fracture was recorded in seconds using a digital chronometer. The number of cycles to failure (NCF) was calculated. The data were analyzed statistically (P < .05). Results: There was a significant difference among the groups (P < .05). The WaveOne Gold Primary showed the greatest mean of NCF (1628 +/- 107), and the WaveOne Primary showed the lowest mean of NCF (1153 +/- 119.2). Conclusions: Within the limitations of this in vitro study, the cyclic fatigue resistance of the WaveOne Gold Primary single-file system was higher than the WaveOne Primary and Reciproc R25 single-file instruments.
机译:简介:这项研究的目的是比较Reciproc R25(VDW,慕尼黑,德国),WaveOne Primary(瑞士登贝利·梅勒费尔(Dentsply Maillefer),Ballaigues)和WaveOne Gold Primary文件(Dentsply Maillefer)的抗循环疲劳性能。方法:本研究包括20台Reciproc R25、20台WaveOne Primary和20台WaveOne-Gold Primary仪器。使用循环疲劳测试设备进行循环疲劳测试,该设备具有人造不锈钢管,该人造不锈钢管的曲率角为60度,曲率半径为5毫米。这些文件被随机分为3组(第1组:Reciproc R25;第2组:WaveOne Primary;以及第3组:WaveOne Gold Primary)。旋转所有仪器直到发生断裂,并使用数字计时器将断裂时间记录在几秒钟内。计算了失效循环数(NCF)。对数据进行统计学分析(P <.05)。结果:各组之间存在显着差异(P <.05)。 WaveOne Gold Primary的NCF平均值最高(1628 +/- 107),WaveOne Primary的NCF平均值最低(1153 +/- 119.2)。结论:在这项体外研究的局限性内,WaveOne Gold初级单锉刀系统的抗循环疲劳性能高于WaveOne初级和Reciproc R25单锉刀。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号