Effects of cleaning, chemical disinfection, and sterilization procedures on the mechanical properties of endodontic instruments.

机译:清洁,化学消毒和灭菌程序对牙髓器械机械性能的影响。

获取原文
获取原文并翻译 | 示例
       

摘要

The aim of this study was to test and compare the values of torsional moment, torsional angular deflection, bending moment, and permanent angular deflection of three designs of root canal files (Unifile, Flexofile, and H-File) before and after cross-infection treatment procedures, according to ANSI/ADA specification no. 28. An increase in value for all mechanical properties tested was observed after the treatment procedures, with the exception of Flexofile wherein a decrease in permanent angular deflection was evident. Unifile showed a decrease in torsional moment and bending moment. The changes in mechanical properties after treatment procedures ranged from 0.1 to 63% from the control groups. Generally, the changes in values observed were insignificant and still well within ANSI/ADA specification no. 28. Thus, they do not have any clinical significance.
机译:这项研究的目的是测试和比较交叉感染前后三种根管锉(Unifile,Flexofile和H锉)的扭转力矩,扭转角挠度,弯曲力矩和永久角挠度的值处理程序,根据ANSI / ADA规范No. 28.在处理程序之后,观察到所有测试的机械性能的值都有所增加,但Flexofile除外,其中永久性角挠度明显降低。 Unifile的扭矩和弯矩减小。对照组治疗后机械性能的变化范围为0.1%到63%。通常,观察到的值变化不明显,仍然在ANSI / ADA规范No.6中。 28.因此,它们没有任何临床意义。