首页> 外文期刊>Journal of endovascular therapy: an official journal of the International Society of Endovascular Specialists >Retroperitoneal aortic hemorrhage caused by penetration of an endovascular stent-graft anchoring barb
【24h】

Retroperitoneal aortic hemorrhage caused by penetration of an endovascular stent-graft anchoring barb

机译:血管内支架-植骨倒刺穿刺导致腹膜后主动脉出血

获取原文
获取原文并翻译 | 示例
       

摘要

Purpose: To report a rare case of acute intraoperative retroperitoneal hemorrhage secondary to aortic penetration by the suprarenal anchoring barb on a stent-graft. Case Report: A 75-year-old patient on dual antiplatelet therapy for coronary stents and lowmolecular-weight heparin for atrial thrombus underwent elective endovascular repair of a 6.7-cm infrarenal abdominal aortic aneurysm. A device with suprarenal fixation and metal anchoring barbs was implanted, and a molding balloon was used that at no time covered the proximal bare metal stents or barbs. In recovery, the patient became tachycardic and hypotensive. After resuscitation, imaging identified an anterior barb penetrating the aortic wall, causing the acute retroperitoneal hemorrhage. A decision to treat conservatively rather than resort to open surgery was difficult but ultimately influenced by the patients' high risk for open surgery. The patient was treated by aggressive reversal of heparin and platelet transfusion, and the bleed settled spontaneously. Conclusion: Major surgery and subsequent morbidity may be avoided by medical management of what would appear to be a surgical problem.
机译:目的:报告少见的支架上移植物经肾上锚钩刺入主动脉穿透继发的急性术中腹膜后出血的病例。病例报告:一名75岁的患者接受了冠状动脉支架和低分子肝素治疗房颤的双重抗血小板治疗,行选择性血管内修复6.7厘米的肾下腹主动脉瘤。植入具有肾上固定和金属锚固倒刺的装置,并使用成型气囊,该气囊在任何时候都不会覆盖近端裸金属支架或倒刺。恢复后,患者心动过速和血压降低。复苏后,影像学检查发现前钩刺穿主动脉壁,引起急性腹膜后出血。保守治疗而不是诉诸于开放手术的决定很困难,但最终受到患者开放手术高风险的影响。通过积极逆转肝素和血小板输注治疗该患者,并自动解决出血。结论:对可能是外科问题的医疗管理可以避免大手术和随后的发病率。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号