Type IIIb Endoleak With the Endurant Stent-Graft

机译:具有耐久性支架移植物的IIIb型内漏

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: To report 2 cases of type IIIb endoleak with the Endurant stent-graft and postulate the cause for the events. Case Report: A type IIIb endoleak was diagnosed at open conversion for a ruptured aneurysm 4 years after implantation of an Endurant stent-graft. In the other case, the endoleak was diagnosed at angiography 4 years after the Endurant stent-graft was implanted; the stent-graft was relined. In both cases the fabric hole was in the body of the stent-graft at the level of the top of the contralateral limb. Conclusion: The cause of the type IIIb endoleaks in these cases was fabric erosion likely due to interaction between the bare metal at the top of the contralateral limb and the fabric of the stent-graft body.
机译:目的:报告2例使用Endurant支架移植物的IIIb型内漏,并推测事件发生的原因。病例报告:在植入Endurant支架移植物4年后,在开放转换时诊断为IIIb型内漏,导致动脉瘤破裂。在另一种情况下,在植入Endurant支架移植物4年后通过血管造影诊断出内漏。覆膜支架移植物。在这两种情况下,织物孔均在对侧肢体顶部水平处的覆膜支架的主体中。结论:在这些情况下,IIIb型内漏的原因是织物腐蚀,这可能是由于对侧肢体顶部的裸露金属与支架移植物的织物之间的相互作用所致。