Suitability of 7 Aortic Stent-Graft Models for MRI-Based Surveillance.

机译:7种基于MRI的主动脉支架移植模型的适用性。

获取原文
获取原文并翻译 | 示例
       

摘要

Purpose: To evaluate the magnetic resonance imaging (MRI) characteristics of commercially available stent-grafts used for abdominal aortic aneurysm repair.Methods: Seven endovascular grafts (AneuRx, Lifepath, Talent, Excluder, Zenith, Quantum LP, and Ancure) were suspended in a water bath containing gadolinium and scanned using a 1.5-T clinical MRI scanner. Two different scan techniques (T(1)-weighted spoiled gradient echo and spin echo) based upon a clinical MRI endograft surveillance protocol were used for each stent-graft. The scans were evaluated for susceptibility artifacts and radiofrequency (RF) shielding and caging artifacts.Results: For most endografts, the lumen and structures surrounding the endograft were well visualized. However, the ferromagnetic properties of the Zenith and Lifepath devices resulted in large susceptibly artifacts that obliterated the endograft lumen as well as adjacent structures. All fully supported grafts showed some amount of signal loss from the graft lumen caused byRF caging. For the Ancure graft, evaluation around the attachment sites might be problematic.Conclusions: An MRI-based surveillance protocol appears to be a viable option for the AneuRx, Talent, Excluder, and Quantum LP stent-grafts.
机译:目的:评估市售用于腹主动脉瘤修复的覆膜支架的磁共振成像(MRI)特性。方法:将七种血管内覆膜支架(AneuRx,Lifepath,Talent,Excluder,Zenith,Quantum LP和Ancure)悬吊于其中。装有g的水浴,并使用1.5-T临床MRI扫描仪进行扫描。对于每个支架移植物,均使用基于临床MRI移植物内窥镜监测方案的两种不同的扫描技术(T(1)加权变质梯度回波和自旋回波)。评估了扫描结果的敏感性伪影以及射频(RF)屏蔽和笼罩伪影。结果:对于大多数内移植物,内移植物周围的内腔和结构均清晰可见。但是,Zenith和Lifepath装置的铁磁特性导致产生大量易感的伪影,这些伪影消除了移植物内腔以及邻近结构。所有完全支持的移植物均显示由RF笼管引起的移植物管腔信号丢失。对于Ancure移植物,附着点周围的评估可能会出现问题。结论:基于MRI的监视协议似乎是AneuRx,Talent,Excluder和Quantum LP支架移植物的可行选择。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号