Do ultrasonic dental scalers interfere with implantable cardioverter defibrillators? An in vivo investigation

机译:超声波洁牙机是否会干扰植入式心脏复律除颤器?体内调查

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Objectives To test the in vivo effects of an ultrasonic dental scaler on various implanted cardioverter defibrillator (ICD) models. Methods 12 consecutive patients with ICDs had continuous both electrocardiogram monitoring and device interrogation to detect interferences during the use of an ultrasonic dental scaler. Results No interferences were detected by any ICD. Evaluation of the electrocardiograms for each patient failed to show any abnormalities in pacing during testing. Conclusion The results of this study suggest that the routinary clinic use of piezoelectric dental scalers do not interfere with the functioning of any of the tested ICDs. Clinical significance Ultrasonic dental scalers have been suspected of electromagnetic interference (EMI) with the normal functioning of ICDs and the use of this type of equipment for patients with these devices has been controversial. This is the first in vivo study to investigate EMI of ICD activity during the operation with ultrasonic dental scaler.
机译:目的测试超声波洁牙机对各种植入式心脏复律除颤器(ICD)模型的体内效果。方法连续12例ICD患者接受连续的心电图监测和设备询问,以检测使用超声波洁牙机期间的干扰。结果任何ICD均未检测到干扰。对每位患者的心电图评估均未显示测试过程中的任何起搏异常。结论这项研究的结果表明,常规的压电洁牙机的临床使用不会干扰任何经过测试的ICD的功能。临床意义人们怀疑超声波洁牙机具有ICD正常功能的电磁干扰(EMI),对于这类设备的患者使用这种类型的设备一直存在争议。这是第一个研究超声洁牙机手术期间ICD活性的EMI的体内研究。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号