首页> 外文期刊>Clinical Chemistry: Journal of the American Association for Clinical Chemists >C-reactive protein (CRP) in neonates: comparing VITROS slide and high-sensitivity CRP methods.
【24h】

C-reactive protein (CRP) in neonates: comparing VITROS slide and high-sensitivity CRP methods.

机译:新生儿C反应蛋白(CRP):比较VITROS玻片和高敏CRP方法。

获取原文
获取原文并翻译 | 示例
           

摘要

To the Editor: C-reactive protein {CRP) is measured on VITROS~R Chemistry Systems (Ortho-Clinical Diagnostics) using a "MicroSlide" method. At our hospitals plasma CRP concentrations <10 mg/L are used as an indicator that it is safe to discontinue antibiotic therapy in neonates with known or suspected sepsis (1). After implementing the neonatal sepsis protocol, we found that the VITROS CRP slide gave results up to 33 mg/L higher in neonatal specimens (Fig. 1, open triangles) than the Behring Nephelom-eter II (BNII, Dade Behring) high-sensitivity CRP (hsCRP) assay, a well-characterized method for CRP measurements (2).
机译:致编辑:C反应蛋白(CRP)在VITROS〜R化学系统(Ortho-Clinical Diagnostics)上使用“ MicroSlide”方法进行测量。在我们的医院中,血浆CRP浓度<10 mg / L被用作指示在已知或怀疑败血症的新生儿中安全停止抗生素治疗的指标(1)。实施新生儿败血症方案后,我们发现VITROS CRP载玻片在新生儿标本中的结果比Behring Nephelometer-ter II(BNII,Dade Behring)高33 mg / L(图1,空心三角)。 CRP(hsCRP)测定法,一种特征明确的CRP测量方法(2)。