摘要

BACKGROUND: High serum concentrations of monoclonal free light chain (FLC) kappa or lambda are markers of plasma cell dyscrasia. METHODS: We developed new, latex-enhanced, specific nephelometric assays based on monoclonal antibodies for the determination of FLC kappa and lambda in serum, EDTA plasma and Li-heparin plasma for use on the Siemens BN systems. RESULTS: Reference ranges were determined from 369 samples: FLC kappa 6.7-22.4 mg/L, FLC lambda 8.3-27.0 mg/L and kappa/lambda ratio 0.31-1.56. Protection from falsely low results due to antigen excess is obtained with a built-in pre-reaction in the assay protocols. Lot-to-lot consistency between three different lots of reagent, calibrators and supplementary reagent lots showed normalized differences <7.5%. The reproducibility of serum samples varied between 4% and 7%. The method comparison with Freelite assays showed normalized differences of 19.7%, 32.7% and 21.7%, respectively, for FLC kappa, lambda and ratio, correlations of 0.94, 0.77 and 0.73, and concordance rates of 99.2%, 94.2% and 95%. CONCLUSIONS: N Latex FLC demonstrates high precision, good lot-to-lot consistency and freedom from a high-dose hook effect. The method comparison between Freelite and the N Latex FLC assays showed good clinical concordance. Further studies need to reveal the clinical value of the new FLC assays.
机译:背景:单克隆游离轻链(FLC)κ或λ的高血清浓度是浆细胞发育不良的标志。方法:我们基于单克隆抗体开发了新的乳胶增强的特异性浊度测定法,用于测定西门子BN系统中使用的血清,EDTA血浆和肝素锂血浆中的FLCκ和λ。结果:从369个样品中确定了参考范围:FLCκ6.7-22.4 mg / L,FLCλ8.3-27.0 mg / L和κ/λ比0.31-1.56。在检测方案中使用内置的预反应可防止因抗原过量而导致错误的低检测结果。三种不同批次的试剂,校准品和补充试剂批次之间的批次间一致性显示标准化差异<7.5%。血清样品的重现性介于4%和7%之间。与Freelite分析的方法比较显示,FLCκ,λ和比率的归一化差异分别为19.7%,32.7%和21.7%,相关性为0.94、0.77和0.73,一致性率为99.2%,94.2%和95%。结论:N Latex FLC证明了高精度,良好的批次间一致性以及不受大剂量钩效应的影响。 Freelite和N Latex FLC分析之间的方法比较显示出良好的临床一致性。进一步的研究需要揭示新的FLC分析的临床价值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号