首页> 外文期刊>Clinical chemistry and laboratory medicine: CCLM >A sensitive quantitative test strip based point-of-care albuminuria screening assay
【24h】

A sensitive quantitative test strip based point-of-care albuminuria screening assay

机译:基于灵敏定量试纸的即时护理蛋白尿筛查方法

获取原文
获取原文并翻译 | 示例

摘要

Background: Chronic kidney disease is a major health problem and the global guidelines require screening of albuminuria. Therefore, affordable and sensitive albuminuria screening tests are needed. We explored the potential of urine strips, generally reported in the ordinal scale, measured on an automatic strip reader for reporting quantitative and sensitive albumin results. Methods: We compared reflectance data of Combur-Test? strips obtained from the Cobas U411 reader (Roche) with albuminuria data from a nephelometer BNII (Siemens) and with protein concentrations from the pyrogallol red method (Modular P, Roche) for 389/328 non-pathologic and pathologic urine samples, respectively. Results: Imprecision of the reflectance signal of the Cobas U411 was measured with commercial control material (Bio-Rad). Inter-run coefficients of variations (CVs) for reflectance for levels 1 and 2 were 1.7/4.9, respectively, and intra-run CVs were 1.8/4.2, respectively. Good agreement was obtained between the albumin concentration of the BNII and the protein strip reflectance data (n389): Y (10000/protein reflectance, 1/)1600.132·X (albuminuria BNII, mg/L)0.0000111·X 2 (albuminuria BNII, mg/L); r 20.921. Lower agreement was found between the protein assay (n328) and the reflectance (r 20.831). A calibration curve was made between 11.5 mg/L and 121.5 mg/L. The limit of blank (LOB) was 44.7 mg/L. Conclusions: The present study demonstrates that reflectance data generated by a test strip reader allows for quantitative analysis of albumin. Although the lower limit of the microalbumin range (30 mg/L) cannot be achieved with the dye-binding method, the results are satisfactory for screening purposes.
机译:背景:慢性肾脏疾病是一个主要的健康问题,全球指南要求对白蛋白尿进行筛查。因此,需要负担得起且敏感的白蛋白尿筛查试验。我们探索了在自动条读取器上测量的尿条的潜力(通常以序数形式报告),用于报告定量和敏感的白蛋白结果。方法:我们比较了Combur-Test的反射率数据?从Cobas U411阅读器(Roche)获得的条带,分别来自比浊仪BNII(西门子)的蛋白尿数据和来自邻苯三酚红法(Modular P,Roche)的蛋白浓度,分别用于389/328非病理性和病理性尿液样本。结果:用商业对照材料(Bio-Rad)测量了Cobas U411反射信号的不精确度。 1级和2级反射的批间变异系数(CV)分别为1.7 / 4.9,批内CV分别为1.8 / 4.2。 BNII的白蛋白浓度与蛋白条反射率数据(n389)吻合良好:Y(10000 /蛋白反射率,1 /)1600.132·X(白蛋白尿BNII,mg / L)0.0000111·X 2(白蛋白尿BNII,毫克/升); r 20.921。在蛋白质测定(n328)和反射率(r 20.831)之间发现较低的一致性。校准曲线在11.5 mg / L和121.5 mg / L之间。空白(LOB)的限值为44.7 mg / L。结论:本研究证明试纸条读取器产生的反射率数据可用于白蛋白的定量分析。尽管用染料结合法无法达到微量白蛋白范围的下限(30 mg / L),但对于筛选目的,结果令人满意。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号