Dextran is used as a plasma expander, and it can interfere with measurement of total serum protein using the biuret assay in strong alkaline solution because it causes turbidity in serum via the formation of an insoluble complex of dextran with copper and tartrate (1). In a recent paper published in this journal, Delanghe et al. demonstrated that the relative total serum protein error due to the addition of dextran 70 was pronounced in various routine chemistry analyzers/reagents, such as the APE/Diagam, Dade, Merck Mega and Roche Integra systems, within the therapeutic range (6 g/L).
机译:右旋糖酐被用作血浆扩展剂,它会在强碱性溶液中使用缩二脲测定法干扰总血清蛋白的测量,因为它会通过形成右旋糖酐与铜和酒石酸盐的不溶性复合物而导致血清混浊(1)。在最近发表在该杂志上的论文中,Delanghe等人。证明在各种常规化学分析仪/试剂(例如APE / Diagam,Dade,Merck Mega和Roche Integra系统)在治疗范围内(6 g / L )。