Commutability of serum protein values: persisting bias among manufacturers using values assigned from the certified reference material 470 (CRM 470) in the United States.

机译:血清蛋白值的可交换性:制造商之间持续存在偏见,使用美国认证参考材料470(CRM 470)分配的值。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Significant among-manufacturer differences in values for serum proteins persist 7 years after the introduction of the international reference material (Certified Reference Material 470; Reference Preparation for Proteins in Human Serum). In some cases, such as transthyretin and C4, the biases actually continue to increase. Further efforts at standardization are needed in order to improve commutability of results among laboratories and are essential if universal reference intervals are to be developed.
机译:制造商之间血清蛋白值的显着差异在引入国际参考材料(认证参考材料470;人血清中蛋白质的参考制备物)后的7年内仍然存在。在某些情况下,例如运甲状腺素蛋白和C4,偏倚实际上会继续增加。为了提高实验室之间结果的可交换性,需要在标准化方面做进一步的努力,如果要建立通用的参考区间,这是必不可少的。