Analytical performance and clinical results of a fully automated MEIA system for brain natriuretic peptide assay: comparison with a point of care testing method.

机译:全自动MEIA系统用于脑钠肽测定的分析性能和临床结果:与即时检验方法的比较。

获取原文
获取原文并翻译 | 示例
           

摘要

The aim of this study was to evaluate the analytical performance of a recently available immunoassay for brain natriuretic peptide (BNP), based on microparticle enzyme immunoassay (MEIA, AxSYM System, Abbott Laboratories), whose analytical characteristics and clinical results were compared with those of a point of care testing (POCT) method (TRIAGE system, Biosite Diagnostics). The within-run and total imprecision of the MEIA system were 18.4% and 19.8% at 21 ng/l, 8.0% and 14.8% at 183 ng/l, and 5.7% and 14.0% at 319 ng/l, respectively. The detection limit of the MEIA system was tested by repeatedly measuring (n=20) the 0 calibrator in four different runs; a mean +3 SD value of 5.6+/-4.8 ng/l (range 1.8-12.6 ng/l) was obtained. A close linear relationship (MEIA= -22.5+/-1.71 POCT method, R=0.950, n=296) was found (BNP concentration: 5-5500 ng/l), with a significant bias (mean difference: 164.8 ng/l, p<0.0001). Mean BNP concentration measured in 94 reference subjects (57 women and 37 men; mean age 43.5+/-14.0 years) was higher with MEIA than POCT, (25.9+/-32.7 ng/l vs. 11.7+/-8.9 ng/l, p<0.0001). The same trend was observed also in 202 cardiac patients (620.6+/-1082.2 ng/l vs. 386.1+/-594.5 ng/l, p<0.0001). Our data suggest that MEIA and POCT have quite similar analytical performance but different clinical results. Then, different reference values, as well as cut-off values, should be taken into account for the clinical use of these two immunoassays.
机译:本研究的目的是评估基于微粒酶免疫测定法(MEIA,AxSYM System,Abbott Laboratories)的脑钠肽(BNP)的最新免疫测定法的分析性能,并将其分析特性和临床结果与即时检验(POCT)方法(TRIAGE系统,Biosite Diagnostics)。 MEIA系统的运行内误差和总不精确度在21 ng / l时分别为18.4%和19.8%,在183 ng / l时分别为8.0%和14.8%,在319 ng / l时分别为5.7%和14.0%。通过在四个不同的运行中重复测量(n = 20)0校准器来测试MEIA系统的检出限。获得的+3 SD平均值为5.6 +/- 4.8 ng / l(范围为1.8-12.6 ng / l)。发现紧密的线性关系(MEIA = -22.5 +/- 1.71 POCT方法,R = 0.950,n = 296)(BNP浓度:5-5500 ng / l),有显着偏差(平均差:164.8 ng / l) ,p <0.0001)。在94名参考受试者(57名女性和37名男性;平均年龄43.5 +/- 14.0岁)中,MEIA的平均BNP浓度高于POCT(25.9 +/- 32.7 ng / l与11.7 +/- 8.9 ng / l)。 ,p <0.0001)。在202名心脏病患者中也观察到了相同的趋势(620.6 +/- 1082.2 ng / l与386.1 +/- 594.5 ng / l,p <0.0001)。我们的数据表明,MEIA和POCT具有相当相似的分析性能,但临床结果却不同。然后,对于这两种免疫测定的临床应用,应考虑不同的参考值以及临界值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号