首页> 外文期刊>Clinical chemistry and laboratory medicine: CCLM >Stability of serum samples and hemolysis interference on the high sensitivity troponin T assay.
【24h】

Stability of serum samples and hemolysis interference on the high sensitivity troponin T assay.

机译:血清样品的稳定性和溶血干扰对高灵敏度肌钙蛋白T测定的影响。

获取原文
获取原文并翻译 | 示例
           

摘要

Studies have shown the negative effect of hemolysis on the troponin T assay (1,2). The troponin T high sensitivity (hs) assay has been recently introduced by Roche Diagnostics (Basel, Switzerland). However, data on stability and hemolysis interference on the new troponin T hs assay are relatively limited. The manufacturer (reagent reference number: 05092744) states that specimens are stable for 24 h at 2-8degC, 12 months at -20degC, and the assay is unaffected by hemolysis (Hb < 0.1 g/dL, sample showing visible signs of hemolysis may cause interference). Therefore, we evaluated the stability in samples at room temperature and the degree of analytical interference from hemolysis with the troponin T hs assay as part of our laboratory method validation.
机译:研究表明溶血对肌钙蛋白T检测有负面影响(1,2)。罗氏诊断公司(瑞士巴塞尔)最近推出了肌钙蛋白T高灵敏度(hs)测定法。然而,关于新肌钙蛋白T hs测定的稳定性和溶血干扰的数据相对有限。制造商(试剂参考编号:05092744)指出,样品在2-8℃下稳定24小时,在-20℃下稳定12个月,该测定不受溶血的影响(Hb <0.1 g / dL,样品显示出明显的溶血迹象造成干扰)。因此,作为我们实验室方法验证的一部分,我们使用肌钙蛋白T hs分析评估了室温下样品的稳定性以及溶血的分析干扰程度。