首页> 外文期刊>Clinical chemistry and laboratory medicine: CCLM >Target analyte quantification by isotope dilution LC-MS/MS directly referring to internal standard concentrations--validation for serum cortisol measurement.
【24h】

Target analyte quantification by isotope dilution LC-MS/MS directly referring to internal standard concentrations--validation for serum cortisol measurement.

机译:通过同位素稀释LC-MS / MS直接参考内标浓度对目标分析物进行定量-验证血清皮质醇的测定。

获取原文
获取原文并翻译 | 示例
           

摘要

Isotope dilution LC-MS/MS methods used in the clinical laboratory typically involve multi-point external calibration in each analytical series. Our aim was to test the hypothesis that determination of target analyte concentrations directly derived from the relation of the target analyte peak area to the peak area of a corresponding stable isotope labelled internal standard compound [direct isotope dilution analysis (DIDA)] may be not inferior to conventional external calibration with respect to accuracy and reproducibility.Quality control samples and human serum pools were analysed in a comparative validation protocol for cortisol as an exemplary analyte by LC-MS/MS. Accuracy and reproducibility were compared between quantification either involving a six-point external calibration function, or a result calculation merely based on peak area ratios of unlabelled and labelled analyte.Both quantification approaches resulted in similar accuracy and reproducibility.For specified analytes, reliable analyte quantification directly derived from the ratio of peak areas of labelled and unlabelled analyte without the need for a time consuming multi-point calibration series is possible. This DIDA approach is of considerable practical importance for the application of LC-MS/MS in the clinical laboratory where short turnaround times often have high priority.
机译:临床实验室中使用的同位素稀释LC-MS / MS方法通常在每个分析系列中涉及多点外部校准。我们的目的是检验以下假设:直接从目标分析物峰面积与相应的稳定同位素标记的内标化合物的峰面积之间的关系得出的目标分析物浓度的确定[直接同位素稀释分析(DIDA)]可能不差相对于传统的外部校准,其准确性和可重复性。通过LC-MS / MS,在比较验证方案中分析了质控样品和人类血清库中皮质醇的示例性分析物。在涉及六点外部校准功能的定量分析或仅基于未标记和标记的分析物的峰面积比进行结果计算之间进行了准确性和可重复性的比较,两种定量方法均具有相似的准确性和可重复性。直接从标记的和未标记的分析物的峰面积之比得出,无需耗时的多点校准系列。这种DIDA方法对于LC-MS / MS在临床实验室中的应用具有相当重要的实践意义,在这些实验室中,短的周转时间通常被优先考虑。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号