Comparison of two point-of-care testing (POCT) devices for fetal lactate during labor

机译:两种分娩期间胎乳的即时检验(POCT)装置的比较

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Background: Point-of-care testing (POCT) of fetal scalp blood lactate is used as an alternative for pH analysis. Lactate measurements have not been standardized and values vary with each device used. The aim of this study was to evaluate the performance of two POCT lactate meters for intrapartum use. Methods: Analytical performance of StatStrip Lactate (Nova Biomedical) and Lactate Pro (Arkray) was evaluated using CLSI EP10. Both POCT meters were compared with our lactate reference method (RapidLab 860; Siemens Healthcare Diagnostics) using fetal scalp and neonatal cord blood. Deming regression analysis was performed. Results: StatStrip Lactate coefficients of variation (CVs) were 5.1 %, 5.0 % and 2.6 % at 0.9, 7.5 and 14.1 mmol/L lactate, respectively. CVs for Lactate Pro were 10.7 %, 5.2 % and 5.7 % at 1.7, 4.1 and 6.4 mmol/L lactate, respectively. Consecutive lactate measurements in 37 fetal scalp and 122 cord blood samples revealed different test characteristics for the two POCT devices. In fetal scalp blood: StatStrip Lactate = 1.13*RapidLab-0.39 (R 2 = 0.907) and Lactate Pro = 0.95*RapidLab-0.03 (R 2 = 0.823). In cord artery blood: StatStrip Lactate = 1.08*RapidLab-0.09 (R 2 = 0.810) and Lactate Pro = 0.72*RapidLab + 0.59 (R 2 = 0.807). Conclusions: Overall performance of both Lactate Pro and StatStrip Lactate was good, with StatStrip Lactate having smallest CVs and closest correlation to our reference method. Both StatStrip Lactate and Lactate Pro can be used as a lactate POCT device for obstetric use.
机译:背景:胎儿头皮血乳酸的即时检测(POCT)被用作pH分析的替代方法。乳酸测量尚未标准化,其值随所使用的每种设备而异。这项研究的目的是评估两个POCT乳酸计在产时使用的性能。方法:使用CLSI EP10评估StatStrip Lactate(Nova Biomedical)和Lactate Pro(Arkray)的分析性能。使用胎儿头皮和新生儿脐带血,将两种POCT测量仪与我们的乳酸参考方法(RapidLab 860; Siemens Healthcare Diagnostics)进行比较。进行了戴明回归分析。结果:乳酸的StatStrip乳酸变异系数(CVs)分别为5.1%,5.0%和2.6%。乳酸Pro的CV在1.7、4.1和6.4 mmol / L乳酸下分别为10.7%,5.2%和5.7%。在37例胎儿头皮和122例脐带血样品中连续进行的乳酸测量显示,这两种POCT设备的测试特性不同。在胎儿头皮血液中:StatStrip乳酸= 1.13 * RapidLab-0.39(R 2 = 0.907)和乳酸Pro = 0.95 * RapidLab-0.03(R 2 = 0.823)。在脐动脉血中:StatStrip乳酸= 1.08 * RapidLab-0.09(R 2 = 0.810),乳酸乳酸= 0.72 * RapidLab + 0.59(R 2 = 0.807)。结论:Lactate Pro和StatStrip Lactate的总体性能都很好,其中StatStrip Lactate的CV最小,与我们的参考方法最相关。 StatStrip Lactate和Lactate Pro均可用作产科用​​乳酸POCT装置。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号