Potential sources of quantification error when retrospectively assessing metacarpal bone loss from historical radiographs by using digital X-ray radiogrammetry: An experimental study

机译:使用数字X射线放射线照相术回顾性评估历史X射线照片的掌骨丢失时量化误差的潜在来源:一项实验研究

获取原文
获取原文并翻译 | 示例
           

摘要

During the past 15yr, digital X-ray radiogrammetry (DXR) has been used to measure metacarpal bone mineral density (BMD). BMD is often measured in existing cohorts where X-ray images were not acquired in accordance with the DXR imaging protocol (DIP). The purpose of the present study was to analyze how deviations from DIP in historical radiographs may affect the reproducibility of DXR-BMD measurements. Cadaver hand phantoms were used to conduct repeat measurements of deviations from DIP with respect to voltage, exposure, lateral displacement, supination, combination of lateral displacement and supination or rotation, extension of the wrist, and edge enhancement. Direct digital radiography (Aristos; Siemens Healthcare, Erlangen, Germany) was used for image acquisition and dxr-online (Sectra, Link?ping, Sweden) for DXR-BMD measurements. The influence of the tested deviations from DIP ranged from 0 to 32.5mg/cm2 (0-6.8%). On repetition with the same specimen, none of the deviations resulted in a within-specimen reproducibility error greater than 2mg/cm2 (0.4%, equivalent to a T-score of 0.042). Among the tested deviations, all except tube voltage had a magnitude greater than the normal measurement noise for the technique and must therefore be considered when planning a study based on historical images.
机译:在过去的15年中,数字X射线放射线照相术(DXR)已用于测量掌骨骨矿物质密度(BMD)。 BMD通常在现有人群中进行测量,而根据DXR成像协议(DIP)不能获取X射线图像。本研究的目的是分析历史X线照片中与DIP的偏差如何影响DXR-BMD测量的可重复性。尸体手模被用于重复测量DIP在电压,暴露,侧向位移,旋后,侧向位移与旋后或旋转,腕部伸展和边缘增强方面的偏差。直接数字射线照相术(Aristos;西门子医疗公司,德国埃尔兰根)被用于图像采集,而dxr在线(Sectra,Linkping,瑞典)被用于DXR-BMD测量。 DIP的测试偏差的影响范围是0到32.5mg / cm2(0-6.8%)。在重复使用同一样品时,没有任何偏差会导致样品内再现性误差大于2mg / cm2(0.4%,等于0.042的T评分)。在测试的偏差中,除管电压之外的所有偏差的幅度均大于该技术的正常测量噪声,因此在基于历史图像进行研究计划时必须考虑这些偏差。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号