Porous graphitic carbon chromatography-tandem mass spectrometry for the study of isoprostanes in human cerebrospinal fluid

机译:多孔石墨碳色谱-串联质谱研究人脑脊液中异前列腺素的含量

获取原文
获取原文并翻译 | 示例
       

摘要

F2-isoprostanes are produced by the non-enzymatic peroxidation of arachidonic acid in membrane phospholipids. This paper describes a new method for the determination of all four classes of F2-isoprostanes in human cerebrospinal fluid (CSF) involving separation on a I mm x 150 turn porous graphitic carbon (PGC) column and detection by triple quadrupole mass spectrometry in negative-ion electrospray mode. The sample pre-treatment consisted of an ultrafiltration step, following which 300 mu l of CSF sample could be injected directly onto a 14 1 mm x 10 mm PGC guard column functioning as a trap for the analytes. The loading solvent was Milli-Q water at 125 mu l/min. After 3 min, the sample was switched into the separation column. The F2-isoprostanes were separated in 20 min using a linear solvent gradient comprising water, methanol, acetonitrile and ammonium hydroxide at a pH of 9.5 and a flow of 50 mu l/min The limit of detection (calculated as 3S/N) was approximately 40 pM (14 pg/ml). The assay was linear within the examined range (18-450 pg/ml), using CSF spiked with iPF2 alpha-III standard (r(2) > 0.995). Repeatability data were calculated for CSF spiked to 90 pg/ml and the relative standard deviation (RSD) obtained was 3% (n = 6). @ 2005 Elsevier B.V. All rights reserved.
机译:F2-异前列腺素是通过膜磷脂中花生四烯酸的非酶促过氧化反应产生的。本文介绍了一种测定人脑脊髓液(CSF)中所有四类F2-异前列腺素的新方法,该方法涉及在1 mm x 150匝多孔石墨碳(PGC)色谱柱上进行分离,并通过三重四极杆质谱在负离子电喷雾模式。样品预处理包括一个超滤步骤,此后可以将300μlCSF样品直接注入到14 1 mm x 10 mm PGC保护柱中,作为分析物的捕集阱。加载溶剂是125毫升/分钟的Milli-Q水。 3分钟后,将样品切换到分离柱中。使用线性溶剂梯度(包括水,甲醇,乙腈和氢氧化铵)在pH值为9.5且流速为50μl/ min的条件下,在20分钟内分离F2-异前列腺烷。检测极限(计算为3S / N)约为40 pM(14 pg / ml)。使用掺有iPF2α-III标准的CSF(r(2)> 0.995),在检测范围内(18-450 pg / ml),测定呈线性。计算加标到90 pg / ml的CSF的重复性数据,获得的相对标准偏差(RSD)为3%(n = 6)。 @ 2005 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号