An ion chromatographic method has been developed for the determination of traces of iodide in saline waters. A Dionex IonPac AS11 anion-exchange column was used with a mobile phase containing sodium chloride in order to remove interferences of the sample matrix in both the chromatographic separation and detection. This matrix-elimination procedure was reinforced by a post-column reaction detection that was both selective and sensitive for iodide and was based on the reaction of iodide with 4,4'-bis(dimethylamino)diphenylmethane in the presence of N-chlorosuccinimide. Detection was carried out at 605 nm. The detection limit for iodide in seawater is at about 0.8 ppb for a 150-mu l injection, and the relative standard deviation at 5 ppb is better than 4%. Bromide is a potential interference, but is well separated from iodide. No interferences from dissolved organic matter in natural samples have been observed. [References: 17]
机译:已经开发出一种离子色谱法,用于测定盐水中的痕量碘。 Dionex IonPac AS11阴离子交换柱与含有氯化钠的流动相一起使用,以消除样品基质在色谱分离和检测中的干扰。柱后反应检测对碘化物具有选择性和敏感性,并基于碘化物与N-氯代琥珀酰亚胺存在下的4,4'-双(二甲基氨基)二苯甲烷反应而得到增强,从而消除了基质清除过程。检测在605nm进行。对于150毫升的注射液,海水中碘化物的检出限约为0.8 ppb,5 ppb的相对标准偏差优于4%。溴化物是一种潜在的干扰物,但与碘化物的分离良好。没有观察到天然样品中溶解有机物的干扰。 [参考:17]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号