K-27 is a bisquaternary asymmetric pyridinium aldoxime-type cholinesterase reactivator of use in the treatment of poisoning with organophosphorous esterase inhibitors. A sensitive, simple and reliable reverse-phase high-performance liquid chromatographic method with electrochemical detection was developed for the measurement of K-27 concentrations in rat brain, cerebrospinal fluid, serum and urine samples. Male Wistar rats were treated intramuscularly with K-27 and the samples were collected 60 min later. Separation was carried out on an octadecyl silica stationary phase and a disodium phosphate solution (pH 3.7) containing citric acid, octane sulphonic acid and acetonitrile served as mobile phase. Measurements were carried out at 30 degrees C at E-ox 0.65 V. The calibration curve was linear through the range of 10-250 ng/mL. Accuracy, precision and the limit of detection calculated were satisfactory according to internationally accepted criteria. Limit of quantitation was 10 ng/mL. The method developed is reliable and sensitive enough for monitoring K-27 levels from different biological samples including as little as 10 mu L of cerebrospinal fluid. The method - with slight modification in the composition of the mobile phase - can be used to measure a wide range of other related pyridinium aldoxime-type cholinesterase reactivators. (c) 2007 Elsevier B.V. All rights reserved.
机译:K-27是一种双季不对称吡啶鎓醛肟型胆碱酯酶活化剂,用于治疗有机磷酸酯酶抑制剂中毒。开发了一种灵敏,简单,可靠的带有电化学检测的反相高效液相色谱方法,用于测量大鼠脑,脑脊液,血清和尿液样品中的K-27浓度。用K-27肌肉注射雄性Wistar大鼠,并在60分钟后收集样品。在十八烷基二氧化硅固定相上进行分离,并将含有柠檬酸,辛烷磺酸和乙腈的磷酸二钠溶液(pH 3.7)用作流动相。在30摄氏度,E-ox 0.65 V下进行测量。校准曲线在10-250 ng / mL范围内呈线性。根据国际公认的标准,计算出的准确度,精密度和检测限均令人满意。定量限为10 ng / mL。所开发的方法可靠且灵敏,足以监测不同生物样品(包括低至10μL脑脊液)中的K-27水平。该方法(在流动相的组成上稍作修改)可用于测量各种其他相关的吡啶鎓醛肟肟型胆碱酯酶活化剂。 (c)2007 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号