Detection of dideuterium in diprotium gas at infinitesimal concentration using a trace reduction detector

机译:使用痕量还原检测器以最小的浓度检测双rot气体中的二氘

获取原文
获取原文并翻译 | 示例

摘要

An analyzer able to detect extremely small concentrations of dideuterium in diprotium gas was developed. To examine its performance, four gas samples of pure diprotium with dideuterium concentrations of 0.25, 0.5, 1, and 5 cm(3)/m(3) were measured using the analyzer, and peak areas and retention times from 18 chromatograms were determined. Results indicated that the peaks were distributed within a relative standard deviation of approximately 15% and a linear correlation existed between peak area and concentration. Retention times were stable, showing a relative standard deviation of 1% with repeated tests and having no correlation with concentration of the four gas samples. (c) 2007 Elsevier B.V. All rights reserved.
机译:开发了一种能够检测双旋ium气体中极低浓度的二氘的分析仪。为了检查其性能,使用分析仪测量了重二uter浓度为0.25、0.5、1和5 cm(3)/ m(3)的四个纯dip的气体样品,并从18个色谱图中确定了峰面积和保留时间。结果表明,峰分布在大约15%的相对标准偏差内,并且峰面积与浓度之间存在线性关系。保留时间稳定,重复测试显示相对标准偏差为1%,并且与四种气体样品的浓度无关。 (c)2007 Elsevier B.V.保留所有权利。