首页> 外文期刊>Journal of chromatography, A: Including electrophoresis and other separation methods >EVALUATION OF DIFFERENT TECHNIQUES FOR PEAK PURITY ASSESSMENT ON A DIODE-ARRAY DETECTOR IN LIQUID CHROMATOGRAPHY
【24h】

EVALUATION OF DIFFERENT TECHNIQUES FOR PEAK PURITY ASSESSMENT ON A DIODE-ARRAY DETECTOR IN LIQUID CHROMATOGRAPHY

机译:液相色谱中二极管阵列检测器峰纯度评估的不同技术评价

获取原文
获取原文并翻译 | 示例
           

摘要

The efficiencies of nine different techniques implemented on a diode-array detector in liquid chromatography for peak homogeneity determination were compared at a concentration level of a drug that can be found in biological samples. Solutions of cefotaxime sodium (analyte) spiked with various amounts of a model impurity (theophylline) were injected under chromatographic conditions giving severe overlap of the peaks (R(s)=0.14). Under the conditions used, the most efficient techniques implemented in the instrument were spectral suppression (ca. 0.5%), derivative spectrum (<1%), followed by spectral overlay, absorbance ratio plot (<5%), multiple-wavelength chromatograms overlay and purity parameter (ca. 5%). The numerical value of the absorbance ratio could also be exploited statistically to detect 0.5% of impurity. Theophylline at levels up to 10% could not be detected using three-dimensional plots, contour diagrams and derivative chromatograms. [References: 10]
机译:比较了在液相色谱中在二极管阵列检测器上实施的用于峰均一性测定的九种不同技术的效率,该效率是在生物样品中可以发现的药物浓度水平上进行的。在色谱条件下注入掺有各种量的模型杂质(茶碱)的头孢噻肟钠(分析物)溶液,使峰严重重叠(R(s)= 0.14)。在所使用的条件下,仪器中最有效的技术是光谱抑制(约0.5%),导数光谱(<1%),然后是光谱叠加,吸光度比图(<5%),多波长色谱图叠加和纯度参数(约5%)。吸光度比的数值也可以统计地用于检测0.5%的杂质。使用三维图,等高线图和导数色谱图无法检测到茶碱含量高达10%。 [参考:10]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号