Serum ferritin concentration is a sensitive measure of body iron stores. The aim of this study was to compare the performance of two commercially available enzyme-linked immunoassays (ELISAs) for serum ferritin: a widely used manual assay kit (Spectro Ferritin MT(R)), and a new fully automated assay (Immulite(R)). We analyzed serum samples from Moroccan school-aged children (n=51) from a rural area with a high prevalence of iron deficiency anemia (IDA). Four replicates of each sample were analyzed using both assays. For the manual method, the interassay repeatability was 24%, 22%, and 11%, and intraassay precision was 18.3%, 9.2%, and 9.1% at increasing serum ferritin concentrations. Using the automated assay, the interassay repeatability was 7%, 6%, and 6%, and intraassay precision was 1.5%, 5.4%, and 5.5% at increasing serum ferritin concentrations. The two assays were well correlated (y=1.16x+1.83; r=0.98). However, the limits of agreement (LOAs) were wide, particularly at low concentrations. A comparison of the assay results with recommended cutoffs for serum ferritin generated sharply different estimates of the prevalence of iron deficiency (ID) in the sample. We conclude that the automated assay has several potential advantages compared to the manual method, including better precision, less operator dependence, and faster sample through-put. J. Clin. Lab. Anal. 19:196-198, 2005. (c) 2005 Wiley-Liss, Inc.
机译:血清铁蛋白浓度是体内铁存储的敏感指标。这项研究的目的是比较两种市售的血清铁蛋白的酶联免疫测定(ELISA)的性能:一种广泛使用的手动测定试剂盒(Spectro Ferritin MT(R))和一种新型的全自动测定(Immulite(R) ))。我们分析了铁缺乏性贫血(IDA)患病率高的农村地区摩洛哥学龄儿童(n = 51)的血清样本。使用两种测定法分析每个样品的四次重复。对于手动方法,在血清铁蛋白浓度增加时,测定间重复性为24%,22%和11%,测定内精确度为18.3%,9.2%和9.1%。使用自动测定法,在血清铁蛋白浓度增加的情况下,测定间重复性为7%,6%和6%,测定内精密度为1.5%,5.4%和5.5%。两种测定法具有良好的相关性(y = 1.16x + 1.83; r = 0.98)。但是,协议限制(LOA)范围很广,尤其是在低浓度下。将测定结果与推荐的血清铁蛋白截留值进行比较,得出样品中铁缺乏症(ID)患病率的估计值截然不同。我们得出的结论是,与手动方法相比,自动测定具有几个潜在的优势,包括更好的精度,更少的操作员依赖性以及更快的样品通量。 J.临床。实验室肛门19:196-198,2005.(c)2005 Wiley-Liss,Inc.

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号