首页> 外文期刊>Journal of Chromatographic Science >Refractive Index Detection Using an Ultraviolet Detector with a Capillary Flow Cell in Preparative SFC
【24h】

Refractive Index Detection Using an Ultraviolet Detector with a Capillary Flow Cell in Preparative SFC

机译:在制备型SFC中使用带有毛细管流动池的紫外检测器进行折射率检测

获取原文
获取原文并翻译 | 示例
           

摘要

An ultraviolet detector with a rapillary flow tell is evaluated for use as a refractive index detector in supercritical fluid chromatography (SFC) using a CO_2 mobile phase. The temperature or pressure control or both of the cell is critical to minimize density-related refractive index changes. Using CO_2, which has a refractive index value that is significantly smaller than that of common organic solvents, leads to higher detector response. Based on a long-term noise level of 2.3 x 10~(-5) refractive index units, the detection limit for n-hexadecyl alcohol at a signal-to-noise ratio of 2 is 30 ug on a 4.6-mm i.d. packed column. This corresponds to the actual detection limit of 30 ng in the capillary cell (at a split ratio of 1:1000). The linear dynamic range of the detector is more than 2 orders of magnitude. This detection method can readily be used for preparative SFC.
机译:评估了带有虹膜流动检测器的紫外线检测器,以用作使用CO_2流动相的超临界流体色谱(SFC)中的折射率检测器。晶胞的温度或压力控制或两者均至关重要,以最小化与密度有关的折射率变化。使用折射率明显小于普通有机溶剂的折射率的CO_2会导致更高的检测器响应。基于2.3 x 10〜(-5)折射率单位的长期噪声水平,信噪比为2时,在4.6 mm内径上正十六烷基醇的检出限为30 ug。填充柱。这对应于毛细管中30 ng的实际检测极限(分流比为1:1000)。检测器的线性动态范围大于2个数量级。这种检测方法可以很容易地用于制备SFC。