A method that uses supercritical fluid chromatography (SFC) with flame-ionization detection (FID) to determine naphthenes (cycloparaffins), paraffins, olefins, and aromatics in gasoline and JP-4 jet fuel is developed. Four packed silica columns in series are used to isolate the aromatics and provide ring number distribution. In addition, the silica columns provide separation of paraffins from the coeluting olefins and naphthenics. The olefins are trapped on a silver-loaded strong cation-exchange silica gel column and removed by back-flushing. Good agreement is found between gas chromatographic hydrocarbon analysis and the SFC method. A relative standard deviation of 0.4% is found for a jet fuel with a naphthene content of 21.6 wt%. At a low naphthenic content of 1.0 wt%, a relative standard deviation of 3.0% is achieved for gasoline. Analysis times are typically 22 min.
机译:开发了一种使用超临界流体色谱(SFC)和火焰电离检测(FID)来确定汽油和JP-4喷气燃料中的环烷(环烷烃),链烷烃,烯烃和芳烃的方法。四个串联的填充硅胶柱用于分离芳烃并提供环数分布。另外,二氧化硅柱可从共洗脱的烯烃和环烷烃中分离出链烷烃。烯烃被捕集在载银的强阳离子交换硅胶柱上,并通过反吹除去。在气相色谱碳氢化合物分析和SFC方法之间发现了很好的一致性。对于环烷含量为21.6重量%的喷气燃料,发现相对标准偏差为0.4%。在1.0 wt%的低环烷烃含量下,汽油的相对标准偏差为3.0%。分析时间通常为22分钟。