首页> 外文期刊>Journal of Chromatographic Science >Determination of fluorine, chlorine and bromine in household products by means of oxygen bomb combustion and ion chromatography
【24h】

Determination of fluorine, chlorine and bromine in household products by means of oxygen bomb combustion and ion chromatography

机译:氧弹燃烧和离子色谱法测定家用产品中的氟,氯和溴

获取原文
获取原文并翻译 | 示例
           

摘要

A method for routine determination of fluorine, chlorine and bromine in household products was developed and validated. In this work, halogen analyses were made based on oxygen bomb combustion followed by ion chromatography (IC). The chromatographic analysis was performed by an IonPac AS19 hydroxide-selective anion-exchange column, a reagent free ion chromatograph eluent generator and an anion self-regenerating suppressor in 10 min. The response was linear (r ≥ 0.9995) in the entire investigated domain. The limit of detection for the halogens was in the range of 2 to 9 × 10~(-3) mg/L and the limit of quantification was lower than 8 mg/Kg with 20 μL of injection volume. The certified reference material of ERM-EC 681k was pretreated using an oxygen bomb combustion procedure to demonstrate the precision of the proposed method. The quantitative analysis results obtained by IC for the target elements were 797 ± 9 mg/Kg chlorine and 786 ± 25 mg/Kg bromine, which were in good agreement with the certified values of 800 ± 4 mg/Kg chlorine, 770 ± 5 mg/Kg bromine for ERM-EC 681k, respectively. This validated method was successfully applied for the analysis of fluorine, chlorine and bromine in household product samples, and the variation of halogen contained among the tested samples was remarkable.
机译:开发并验证了一种日常测定家用产品中氟,氯和溴的方法。在这项工作中,基于氧弹燃烧然后进行离子色谱法(IC)进行卤素分析。色谱分析是在10分钟内通过IonPac AS19氢氧化物选择性阴离子交换柱,无试剂离子色谱洗脱液产生器和阴离子自生抑制剂进行的。在整个研究范围内,响应是线性的(r≥0.9995)。进样量为20μL时,卤素的检出限为2至9×10〜(-3)mg / L,定量限低于8 mg / Kg。使用氧弹燃烧程序对ERM-EC 681k的认证参考材料进行了预处理,以证明所提出方法的精度。离子色谱对目标元素的定量分析结果为797±9 mg / Kg氯和786±25 mg / Kg溴,与800±4 mg / Kg氯,770±5 mg的认证值非常一致/ Kg溴分别用于ERM-EC 681k。该验证方法成功地用于家庭产品样品中的氟,氯和溴的分析,被测样品中卤素的含量变化显着。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com