A subtle artifact of patchy hypodensities in computed tomography images of the head mimicked acute or subacute cerebral infarct. The cause of the artifact was air in the oil of the x-ray tube. The artifact manifested only when the acquisition parameters included a rotation time of 0.5 second and a gantry tilt angle of 11 to 20 degrees. Routine quality control testing did not detect nonuniformities in the water phantom.
机译:头部计算机断层扫描图像中微妙的斑块状低密度伪影模仿了急性或亚急性脑梗塞。造成伪影的原因是X射线管的油中有空气。仅当采集参数包括0.5秒的旋转时间和11至20度的龙门架倾斜角时才出现伪影。常规质量控制测试未检测到水体模中的不均匀性。