首页> 外文期刊>Journal of computer assisted tomography >Quantitative and qualitative comparison of standard-dose and low-dose pediatric head computed tomography: A retrospective study assessing the effect of adaptive statistical iterative reconstruction
【24h】

Quantitative and qualitative comparison of standard-dose and low-dose pediatric head computed tomography: A retrospective study assessing the effect of adaptive statistical iterative reconstruction

机译:标准剂量和小剂量儿科头颅计算机断层扫描的定量和定性比较:一项回顾性研究,评估自适应统计迭代重建的效果

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVES: Our aim was to assess the effect of adaptive statistical iterative reconstruction (ASIR) on pediatric head computed tomography (CT) examinations. METHODS: We retrospectively reviewed 305 pediatric head CT examinations. The study population consisted of standard dose (STD, n = 152) examinations reconstructed with filtered back projection and low dose (LD, n = 153) examinations reconstructed with 30% (LD30) and 0% (LD0) ASIR. We compared groups by means of radiation dose, noise measures, and visual grading. Student t test, 1-way analysis of variance test, and Mann-Whitney U test were used for statistical analysis. RESULTS: The dose in the LD30 group was significantly lower (29%) than that in the STD group (P < 0.001). The noise in the white matter (P < 0.001), SNR (P < 0.001), and subjective image noise (P = 0.044) was significantly better in the STD group than those in the LD30 group. There was no significant difference between LD30 and STD groups in the sharpness (P = 0.141), diagnostic acceptability (P = 0.079), and artifacts (P = 0.750) and contrast-to-noise ratio (P = 0.718). CONCLUSIONS: In conclusion, we found that a blend of 30% ASIR in a 16-slice multidetector CT produces diagnostically acceptable pediatric head CT examinations with a 29% less dose.
机译:目的:我们的目的是评估自适应统计迭代重建(ASIR)对小儿头部计算机断层扫描(CT)检查的效果。方法:我们回顾了305例儿科头颅CT检查。研究人群包括标准剂量(STD,n = 152)检查,该检查通过过滤反投影重建,低剂量(LD,n = 153)检查通过30%(LD30)和0%(LD0)ASIR重建。我们通过辐射剂量,噪声测量和视觉等级对组进行了比较。统计分析使用学生t检验,方差检验的1向分析和Mann-Whitney U检验。结果:LD30组的剂量显着低于STD组(29%)(P <0.001)。 STD组的白质噪声(P <0.001),SNR(P <0.001)和主观图像噪声(P = 0.044)明显优于LD30组。 LD30和STD组之间的清晰度(P = 0.141),诊断可接受性(P = 0.079),伪影(P = 0.750)和对比噪声比(P = 0.718)之间没有显着差异。结论:总之,我们发现在16层多层螺旋CT中将30%的ASIR掺入可产生诊断上可接受的小儿头颅CT检查,且剂量减少29%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号