Validation of dual-source single-tube reconstruction as a method to obtain half-dose images to evaluate radiation dose and noise reduction: Phantom and human assessment using CT colonography and sinogram-affirmed iterative reconstruction (SAFIRE)

机译:验证双源单管重建作为获取半剂量图像以评估辐射剂量和降噪的方法:使用CT结肠造影和正弦图确认的迭代重建(SAFIRE)进行幻影和人体评估

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVE: To evaluate a method for obtaining half-dose CT images for observer studies evaluating lower-dose CT. METHODS: Phantoms of varying sizes were scanned at multiple tube potentials using dose-matched dual-source (DS) and single-source (SS) protocols. Images from single-tube reconstruction of DS data were compared with SS images acquired at half-original CTDIvol. Thirty patients underwent supine SS and dose-matched prone DS CT colonography (CTC). Half-dose prone images were reconstructed with sinogram-affirmed iterative reconstruction (SAFIRE). Two radiologists scored image quality on 2-dimensional (2D) and 3D images. RESULTS: Image noise was similar between half-dose SS images and DS images reconstructed from one tube only with tube potential of 120 kV or more for phantoms 40 cm or smaller (P < 0.05). For both readers, the patients' CTC image quality scores were more than 84% concordant between SS or DS CTC images, and half-dose-prone CTC images with SAFIRE had 84% or more concordance with routine-dose CTC except for 3D image noise. CONCLUSIONS: In appropriately sized patients, DS acquisition with single-tube reconstruction can create half-dose images, permitting comparison to full-dose images. For CTC, there is comparable image quality for colonic evaluation between full-dose and half-dose images reconstructed with SAFIRE.
机译:目的:评估获得半剂量CT图像的方法,以供观察者研究评估低剂量CT。方法:使用剂量匹配的双源(DS)和单源(SS)方案在多管电势下扫描大小不同的幻影。将DS数据的单管重建图像与半原始CTDIvol采集的SS图像进行比较。 30名患者接受仰卧位SS和剂量匹配的俯卧DS CT结肠造影(CTC)。半剂量俯卧图像用正弦图确认的迭代重建(SAFIRE)重建。两位放射科医生对2维(2D)和3D图像的图像质量进行了评分。结果:半剂量的SS图像和DS图像之间的图像噪声相似,而对于40 cm或更小的幻像,仅从一支管电位为120 kV或更高的管重建的DS图像(P <0.05)。对于这两个读者,患者的CTC图像质量得分在SS或DS CTC图像之间的一致性超过84%,具有半剂量的SAFIRE的CTC图像与常规剂量的CTC的一致性达到84%或更高,除了3D图像噪声。结论:在适当大小的患者中,通过单管重建进行DS采集可以创建半剂量图像,从而可以与全剂量图像进行比较。对于CTC,在用SAFIRE重建的全剂量和半剂量图像之间,结肠评估的图像质量相当。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号