OBJECTIVES: To study the effect of motion velocity on image quality to determine the requirements for 4-dimensional (4D; ie, 3D + time) musculoskeletal computed tomographic (CT) imaging. MATERIALS AND METHODS: A phantom with resolution targets in both axial (x-y) and coronal (x-z) planes was attached to a motion device and scanned with 64-slice CT using a retrospectively gated CT protocol with pitch values of 0.1 and 0.2. Data were acquired with the phantom at rest and while moving periodically along the x axis at several velocities. Spatial resolution and motion artifacts were assessed both for the axial and coronal targets. RESULTS: A linear relationship was found between motion artifact severity and phantom velocity. Spatial resolution was better preserved in the coronal target. However, coronal images displayed banding artifacts, with band displacements being linearly related to motion velocity. CONCLUSIONS: The 4D CT imaging of periodically moving objects with velocities up to 20 mm/s is feasible using a pitch value of 0.1 and a motion frequency of 30 cycles per minute.
机译:目的:研究运动速度对图像质量的影响,以确定4维(4D;即3D +时间)肌肉骨骼计算机断层扫描(CT)成像的要求。材料与方法:将具有在轴向(x-y)和冠状(x-z)平面上均具有分辨率目标的模型连接到运动设备,并使用回顾性门控CT协议以间距值为0.1和0.2的64层CT进行扫描。数据是在模型处于静止状态并同时沿多个速度沿x轴定期移动时获取的。评估了轴向和冠状目标的空间分辨率和运动伪影。结果:运动伪影严重程度与幻影速度之间存在线性关系。空间分辨率更好地保留在冠状靶中。然而,冠状图像显示出带状伪影,其中带位移与运动速度线性相关。结论:使用0.1的螺距值和30分钟/分钟的运动频率,对速度高达20 mm / s的周期性运动物体进行4D CT成像是可行的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号