Heart rate adaptive optimization of spatial and temporal resolution for electrocardiogram-gated multislice spiral CT of the heart.

机译:心电门控多层螺旋CT的时空分辨率心率自适应优化。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: We introduce a reconstruction method for electrocardiogram (ECG)-gated multislice spiral computed tomography (CT) examinations of the heart [adaptive cardio volume (ACV) reconstruction]. It is evaluated for a four-slice CT system (Siemens Somatom VolumeZoom). METHOD: State-of-the-art reconstruction techniques for ECG-gated multislice spiral CT use scan data from N consecutive heart cycles for image reconstruction. With increased N, the temporal resolution improves up to t rot /(2 N ) ( t rot is the 360 degrees rotation time of the scanner) but at the expense of insufficient volume coverage or loss of longitudinal resolution, especially at low heart rates. With the ACV technique, the number N of consecutive heart cycles used for image reconstruction is automatically adapted to the momentary heart rate of the patient, ranging from N = 1 at very low heart rates up to N = 3 at high heart rates, to maintain both high z resolution (reconstructed slice width close to the collimated slice width) and adequate temporal resolution. We evaluated slice sensitivity profiles and investigated 10 patients with different heart rates ranging from 55 to 110 beats/min for CT angiography (CTA) studies of the coronary arteries and compared the results with those from a reconstruction with fixed N ( N = 1 and N = 2). Axial images as well as multiplanar reformations were used for an evaluation of image quality. RESULTS: With the ACV approach, the complete heart may be scanned at 1 mm slice width within 25-35 s. A narrow slice sensitivity profile (full width at half-maximum of approximately 1.3 mm) is maintained for all heart rates. Diagnostic results can be obtained for heart rates up to about 95 beats/min by individual patient optimization of the ECG gating parameters. Improved temporal resolution at the expense of reduced longitudinal resolution may degrade the image quality of CTA studies at low heart rates by blurring plaques and stenoses. CONCLUSION: The results indicate the potential of the ACV reconstruction technique for high-resolution coronary CTA in a wide range of heart rates.
机译:目的:我们介绍一种用于心脏的心电图(ECG)门控多层螺旋CT(CT)检查的重建方法[自适应心脏容积(ACV)重建]。对四层CT系统(Siemens Somatom VolumeZoom)进行了评估。方法:ECG门控多层螺旋CT的最新重建技术使用来自N个连续心动周期的扫描数据进行图像重建。随着N的增加,时间分辨率提高到t rot /(2 N)(t rot是扫描仪的360度旋转时间),但代价是体积覆盖率不足或纵向分辨率降低,尤其是在低心率时。通过ACV技术,用于图像重建的N个连续心动周期数N会自动适应患者的瞬时心率,范围从极低心率下的N = 1到高心率下的N = 3高z分辨率(重构的切片宽度接近准直切片宽度)和足够的时间分辨率。我们评估了切片的敏感性曲线,并调查了10名不同心率的患者,这些患者的心率范围从55至110次/分钟,用于冠状动脉的CT血管造影(CTA)研究,并将结果与​​固定N(N = 1和N = 2)。轴向图像以及多平面重建用于评估图像质量。结果:使用ACV方法,可以在25-35 s内以1 mm的切片宽度扫描整个心脏。对于所有心率,都保持较窄的切片灵敏度曲线(半峰全宽约为1.3毫米)。通过对患者的ECG门控参数进行优化,可以获得高达95次/分钟的心率的诊断结果。以降低的纵向分辨率为代价的改进时间分辨率可能会降低斑块和狭窄,从而降低低心率时CTA研究的图像质量。结论:结果表明ACV重建技术在广泛的心率范围内可用于高分辨率冠状动脉CTA。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号