首页> 外文期刊>Journal of cataract and refractive surgery >Comparison of actual vacuum pressures at the end of 3 phacoemulsification tips in swine eyes.
【24h】

Comparison of actual vacuum pressures at the end of 3 phacoemulsification tips in swine eyes.

机译:猪眼3个超声乳化吸头末端实际真空压力的比较。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To compare the actual vacuum pressure at the end of an aspiration bypass system (ABS) phaco tip and a conventional tip in conditions similar to those during human cataract surgery. SETTING: Kangbuk Samsung Hospital, Seoul, Korea. METHODS: A 4.0 mm diameter rubber eraser ball was inserted in a phacoemulsified swine eye and engaged to a phaco tip. With a vacuum pressure of 300 mm Hg, the eraser ball was pulled in the opposite direction with the phaco tip. The pulling power separating the eraser ball from the phaco tip was measured using the volume of water dripped into a container connected to the eraser ball. Measurements were taken with an Infiniti ABS tip and 2 conventional tips: Sovereign (conventional tip S) and Infiniti (conventional tip I). Measurements were taken alternatively at random in the same eye. RESULTS: The mean pulling power was 17.35 mL +/- 5.01 (SD) with conventional tip S and 16.85 +/- 4.45 mL with conventional tip I. The mean pulling power of the ABS tip was 13.35 +/- 4.32 mL, which was a mean 20.8% lower than that of conventional tip I (P = .016) and 23.1% lower than that of conventional tip S (P = .010). CONCLUSIONS: The actual vacuum pressure of the phaco tip was significantly influenced by the presence of the ABS hole. To compensate for the decrease in actual vacuum pressure, surgeons should be aware of the characteristics of the ABS tip being used and change the standard settings accordingly.
机译:目的:在类似于人白内障手术的条件下,比较抽吸旁路系统(ABS)超声乳化末梢和常规超声末梢的实际真空压力。地点:韩国首尔江北三星医院。方法:将直径为4.0 mm的橡皮球插入超声乳化猪眼,并与超声乳头接合。在300 mm Hg的真空压力下,橡皮球用phaco吸头朝相反的方向拉动。使用滴入连接到橡皮擦球的容器中的水量来测量将橡皮擦球与超声乳头分开的拉力。使用英菲尼迪ABS吸头和2个常规吸头进行测量:主权(常规吸头S)和英菲尼迪(常规吸头I)。交替地在同一只眼中进行测量。结果:常规吸头S的平均拉力为17.35 mL +/- 5.01(SD),常规吸头I的平均拉力为16.85 +/- 4.45mL。ABS吸头的平均拉力为13.35 +/- 4.32 mL,为平均比传统刀头I低20.8%(P = 0.016),比传统刀头S低23.1%(P = 0.0110)。结论:ABS孔的存在显着影响了超声乳头的实际真空压力。为了补偿实际真空压力的降低,外科医生应了解所使用的ABS尖端的特性,并相应地更改标准设置。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号