Visual function after implantation of an aspheric bifocal intraocular lens.

机译:非球面双焦点人工晶状体植入后的视觉功能。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

PURPOSE: To evaluate distance, intermediate, and near vision after bifocal aspheric AcrySof ReSTOR SN6AD1 intraocular lens (IOL) implantation. SETTING: Fernandez-Vega Ophthalmological Institute, Oviedo, Spain. METHODS: Binocular distance best corrected visual acuity (BCVA) (4.0 m), best distance-corrected near (40.0 cm) and intermediate (70.0, 60.0, 50.0 cm) acuities, and contrast sensitivity under photopic (85.0 candelas [cd]/m(2)) and mesopic (3.0 cd/m(2)) conditions with and without glare were measured postoperatively. A patient-satisfaction and visual phenomena questionnaire was administered. RESULTS: Six months postoperatively, the mean acuities (logMAR) were binocular BCVA, -0.064 +/- 0.049 (SD); best corrected near, -0.041 +/- 0.061; best corrected intermediate, 0.147 +/- 0.130 (70.0 cm), 0.036 +/- 0.133 (60.0 cm), and -0.126 +/- 0.077 (50.0 cm). The BCVA and best corrected near acuity were 20/25 or better in all patients. Through-focus binocular logMAR acuities had 2 peaks at the expected far focus and near focus (0.00 diopter [D] and -2.40 D, respectively), with somewhat reduced acuity (20/32) at intermediate distances. Photopic contrast sensitivity was within the standard normal range. Mesopic contrast sensitivity was lower than under photopic conditions, particularly at higher spatial frequencies. Patients reported good performance for distance, near, and intermediate visual tasks with no or minimal difficulty with specific tasks. No patient reported severe visual phenomena; halos and glare were rated as none to moderate. CONCLUSIONS: The aspheric IOL provided good functional vision at far and near, with lower performance at intermediate distance; good photopic contrast sensitivity and lower mesopic performance; and a low incidence of visual disturbances.
机译:目的:评估双焦点非球面AcrySof ReSTOR SN6AD1人工晶状体(IOL)植入后的远距,中距和近视力。地点:西班牙奥维耶多的费尔南德斯·维加眼科研究所。方法:双眼距离最佳矫正视力(BCVA)(4.0 m),近距离(40.0 cm)和中间(70.0、60.0、50.0 cm)视力最佳矫正,在明视下(85.0坎德拉[cd] / m)的对比敏感度(2))和中视(3.0 cd / m(2))条件在有无眩光的情况下进行了术后测量。进行了患者满意度和视觉现象调查表。结果:术后六个月的平均视力(logMAR)为双眼BCVA,-0.064 +/- 0.049(SD);最佳校正附近-0.041 +/- 0.061;最佳校正中间体0.147 +/- 0.130(70.0 cm),0.036 +/- 0.133(60.0 cm)和-0.126 +/- 0.077(50.0 cm)。所有患者的BCVA和最佳矫正近视度数均为20/25或更高。通过焦点的双眼logMAR视力在预期的远焦点和近焦点处分别有2个峰值(分别为0.00屈光度[D]和-2.40 D),在中间距离处视力有所降低(20/32)。视觉对比度敏感度在标准正常范围内。中观对比敏感度低于在明暗条件下,特别是在较高的空间频率下。患者报告在远距离,近距离和中等视觉任务上表现良好,对特定任务没有或仅有最小难度。没有患者报告严重的视觉现象;光环和眩光被定为中度。结论:非球面人工晶状体在远近提供了良好的功能性视觉,在中距离具有较低的性能;良好的明暗对比敏感度和较低的中观性能;视觉障碍的发生率低。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号