Comparative performance of the Zyoptix XP and Hansatome zero-compression microkeratomes.

机译:Zyoptix XP和Hansatome零压缩微角膜刀的比较性能。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

PURPOSE: To compare flap parameters produced by the Zyoptix XP and Hansatome microkeratomes (both Bausch & Lomb) and to evaluate preoperative variables contributing to flap thickness variation. SETTING: Private practice, St. Louis, Missouri, USA. METHODS: Flap dimensions in 75 eyes that prospectively had laser in situ keratomileusis using the Zyoptix XP microkeratome were compared with a historical control sample of 75 eyes treated with the Hansatome microkeratome. The 2 groups were matched for mean keratometry, central corneal pachymetry, spherical equivalent (SE), age, and microkeratome head size and suction ring diameter. RESULTS: The mean ultrasound-measured flap thickness was 126.54 microm +/- 14.6 (SD) and 143.74 +/- 15.0 microm for the 120 microm and 140 microm Zyoptix XP heads, respectively, and 128.90 +/- 20.4 microm and 143.32 +/- 21.0 microm for the 160 microm and 180 microm Hansatome heads, respectively. The standard deviation in flap thickness was smaller and statistically significant for the Zyoptix XP (+/-14.8 microm) versus the Hansatome (+/-20.7 microm) (P = .0039, F test). Preoperative pachymetry and SE (P<.001) accounted for 20% of the variability in measured flap thickness using the Hansatome. For the Zyoptix XP, preoperative pachymetry was statistically significant in explaining the variation in flap thickness with the 120 microm head (P = .02) but not with the 140 microm head. Variation in flap thickness from either Zyoptix XP head was not statistically related to the preoperative SE. CONCLUSION: Although the 2 microkeratomes produced flaps of similar mean thickness, the Zyoptix XP showed significantly less variation in flap thickness than the Hansatome, was less affected by measurable preoperative variables such as SE, and was closer to nominal labeling.
机译:目的:比较Zyoptix XP和Hansatome微型角膜刀(包括Bausch&Lomb)产生的皮瓣参数,并评估有助于皮瓣厚度变化的术前变量。地点:美国密苏里州圣路易斯的私人诊所。方法:使用Zyoptix XP微型角膜刀,对75眼预期具有激光原位角膜磨镶术的眼的皮瓣尺寸与经Hansatome微型角膜刀治疗的75眼历史对照样本进行比较。两组患者的平均角膜曲率,中央角膜厚度,球当量(SE),年龄,微角膜刀头大小和吸引环直径相匹配。结果:120微米和140微米Zyoptix XP头的超声测量皮瓣平均厚度分别为126.54微米+/- 14.6(SD)和143.74 +/- 15.0微米,分别为128.90 +/- 20.4微米和143.32 + / -160微米和180微米Hansatome喷头分别为21.0微米。与Hansatome(+/- 20.7微米)相比,Zyoptix XP(+/- 14.8微米)的襟翼厚度的标准偏差更小且具有统计学意义(P = .0039,F检验)。术前测厚法和SE(P <.001)占Hansatome测量皮瓣厚度变化的20%。对于Zyoptix XP,术前测厚法在解释120微米头(P = .02)而不是140微米头的瓣厚度变化方面具有统计学意义。 Zyoptix XP任一头的皮瓣厚度变化与术前SE均无统计学关系。结论:尽管2个微角膜刀产生的皮瓣平均厚度相似,但Zyoptix XP的皮瓣厚度变化远小于Hansatome,受术前可测量变量如SE的影响较小,并且更接近标称标记。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号