首页> 外文期刊>Journal of cataract and refractive surgery >Anterior haptic flexing and in-the-bag subluxation of an accommodating intraocular lens due to excessive capsular bag contraction
【24h】

Anterior haptic flexing and in-the-bag subluxation of an accommodating intraocular lens due to excessive capsular bag contraction

机译:由于囊袋过度收缩而导致容纳性人工晶状体的前触觉屈曲和袋内半脱位

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

We describe the case of a patient who had cataract surgery with implantation of the hydrophilic acrylic Tetraflex accommodating intraocular lens (IOL), with subsequent development of capsulorhexis phimosis and in-the-bag IOL subluxation. Contraction of the capsular bag secondary to fibrosis resulted in significant anterior flexing of the lens haptic component. Explantation of the IOL capsular bag complex was required 7 years after implantation. Histopathologic analysis demonstrated multiple areas of thick anterior subcapsular fibrosis. Pseudoexfoliative material was present throughout the surface of the lens capsule. Intraocular lenses manufactured from hydrophilic acrylic material are highly flexible and may be more susceptible to capsule contraction, even in the absence of predisposing ocular and systemic conditions. This case highlights the importance of developing guidelines regarding patient screening and selection for the appropriate use of accommodating and other highly flexible IOLs.
机译:我们描述了一个患者的案例,该患者进行了白内障手术,植入了亲水性丙烯酸Tetraflex适应性人工晶状体(IOL),随后发生了撕囊性包茎和袋内IOL半脱位。继发于纤维化的囊袋收缩导致晶状体触觉组件明显向前弯曲。植入后7年,需要植入IOL囊袋复合体。组织病理学分析显示前囊膜下纤维化的多个区域。假性剥脱材料存在于整个晶状体囊表面。由亲水性丙烯酸材料制成的人工晶状体具有很高的柔韧性,即使在没有易患眼病和全身疾病的情况下,也可能更容易发生胶囊收缩。该案例强调了制定有关患者筛查和选择指南的重要性,以适当使用容纳性和其他高度灵活的IOL。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号