Performance comparison of the Alcon Legacy 20000 1.1 mm TurboSonics and 0.9 mm Aspiration Bypass System tips.

机译:Alcon Legacy 20000 1.1 mm TurboSonics和0.9 mm Aspiration Bypass系统吸头的性能比较。

获取原文
获取原文并翻译 | 示例

摘要

PURPOSE: To compare the efficiencies of the operative procedures and the reductions in postoperative corneal endothelial cell density (ECD) after cataract surgery using 2 different phacoemulsification tips and their associated cassettes and tubing. SETTING: Wolfe Clinic, Marshalltown, Iowa, USA. METHODS: A randomized prospective study of 100 consecutive cases was conducted. All patients were adults having phacoemulsification using a hybrid in situ fracture technique. All cases were performed by 1 surgeon using the Alcon Legacy 20000 phacoemulsification machine. One of 2 45 degree phacoemulsification tips and associated cassettes and tubing were used: the 1.1 mm TurboSonics with normal tubing (n = 47) or the 0.9 mm Aspiration Bypass System (ABS) tip with high-vacuum cassettes and tubing (n = 53). Measurements at the time of surgery included metered phacoemulsification time, percentage power used, total time spent in the phacoemulsification process, and milliliters of balanced salt solution (BSS) used. Corneal ECD was determined preoperatively and 10 weeks postoperatively. An independent statistician performed 2-sample t tests or Wilcoxon rank sum tests on the data. RESULTS: There were no cases of anterior radial capsule tear, posterior capsule tear, vitreous loss, iris aspiration, incision leak, or suture closure. There was no correlation between amount of corneal ECD loss and any operative variable measured. Similar measurements for the 1.1 mm TurboSonics and the 0.9 mm ABS included, respectively, ECD loss 7.34% and 8.22%; metered phacoemulsification time 1.29 and 1.22 minutes; total time of the phacoemulsification process 113 and 105 seconds. There were statistically significant differences between the 1.1 mm TurboSonics tip and 0.9 mm ABS tip performance in mean power percentage overall (52% versus 48%) and overall BSS volume used (85 versus 76 mL). CONCLUSIONS: The 0.9 mm ABS phacoemulsification tip used with the high-vacuum cassette and tubing provided the physical advantages of tip-size reduction while requiring similar ultrasonic power and BSS volumes as the 1.1 mm TurboSonics tip with standard cassette and tubing. Together, they provide a wider available dynamic range in which to integrate ultrasonic power, higher vacuum, and higher aspiration flow rates to improve the efficiency and ease of use than the standard 1.1 mm TurboSonics cataract operation.
机译:目的:比较白内障手术后使用两种不同的超声乳化术尖端及其相关的盒和导管,以比较手术程序的效率和术后角膜内皮细胞密度(ECD)的降低。地点:美国爱荷华州马歇尔敦的沃尔夫诊所。方法:对100例连续病例进行了一项随机前瞻性研究。所有患者均为使用混合原位骨折技术进行超声乳化术的成年人。所有病例均由1位外科医生使用Alcon Legacy 20000超声乳化仪进行。使用了2个45度超声乳化吸头以及相关的盒和管中的一个:带有普通管的1.1 mm TurboSonics(n = 47)或带高真空盒和管的0.9 mm抽吸旁路系统(ABS)吸头(n = 53) 。手术时的测量包括计量的超声乳化时间,使用的功率百分比,在超声乳化过程中花费的总时间以及所用的平衡盐溶液(BSS)的毫升数。术前和术后10周确定角膜ECD。独立统计学家对数据进行了2样本t检验或Wilcoxon秩和检验。结果:无radial前囊撕裂,后囊撕裂,玻璃体丢失,虹膜抽吸,切口漏出或缝合线闭合的病例。角膜ECD丢失量与任何手术变量之间均无相关性。对于1.1毫米TurboSonics和0.9毫米ABS的相似测量分别包括ECD损失7.34%和8.22%;计量超声乳化时间1.29和1.22分钟;超声乳化过程的总时间为113和105秒。在1.1毫米TurboSonics吸头和0.9毫米ABS吸头的性能之间,在总体平均功率百分比(52%对48%)和所使用的总BSS体积(85对76 mL)之间存在统计学差异。结论:与高真空管和管一起使用的0.9 mm ABS超声乳化吸头提供了减小管头尺寸的物理优势,同时需要与标准管和管的1.1 mm TurboSonics管头相似的超声功率和BSS体积。它们共同提供了更宽的动态范围,可在其中集成超声功率,更高的真空度和更高的抽吸流量,从而比标准的1.1毫米TurboSonics白内障手术更有效地使用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号