首页> 外文期刊>Journal of Clinical Neurophysiology >Success rate of motor evoked potentials for intraoperative neurophysiologic monitoring: effects of age, lesion location, and preoperative neurologic deficits.
【24h】

Success rate of motor evoked potentials for intraoperative neurophysiologic monitoring: effects of age, lesion location, and preoperative neurologic deficits.

机译:术中神经生理监测运动诱发电位的成功率:年龄,病变部位和术前神经功能缺损的影响。

获取原文
获取原文并翻译 | 示例
           

摘要

Transcranial electrical stimulation with myogenic motor evoked potential (MEP) recording was used for intraoperative neurophysiologic monitoring in 341 consecutive "high-risk" neurosurgical or orthopedic procedures. Overall, the success rate for establishing reliable MEP response was 94.8% for upper extremities and 66.6% for lower extremities. The rate was only 39.1% for lower extremities in patients with preoperative motor deficit and up to 81% in neurologically intact adults. Further analysis demonstrated that extremes of age or the presence of a lesion in the spinal cord and motor deficit contributed to failure in obtaining reliable MEPs.
机译:经颅电刺激与肌运动诱发电位(MEP)记录用于341个连续的“高危”神经外科或骨科手术中的术中神经生理监测。总体而言,建立可靠的MEP反应的成功率为上肢94.8%,下肢66.6%。术前运动缺陷患者下肢的比率仅为39.1%,而神经功能完整的成年人的比率高达81%。进一步的分析表明,年龄过大或脊髓中存在病变以及运动功能障碍均导致无法获得可靠的MEP。