Sensitive determination of irinotecan (CPT-11) and its active metabolitel SN-38 in human serum using liquid chromatography-electrospray mass spectrometry

机译:液相色谱-电喷雾质谱法灵敏测定人血清中的伊立替康(CPT-11)及其活性代谢物SN-38

获取原文
获取原文并翻译 | 示例
           

摘要

A couple of sensitive and accurate liquid chromatography-electrospray mass spectrometry (LC-ES-MS) methods for the determination of the total forms of irinotecan and its active metabolite SN-38 in human serum, using the same chomatographic and detection conditions, is presented. Both used camptothecin as internal standard (I.S.).The sample pretreatment for irinotecn involved a simple protein precipitation with acetonitrile, whereas a liquid-liuqid extraction was necessary for SN-38. A Symmetry C_(18),3.5 #mu#m (150×1 mm I.D) reversed-phase column was used for the chromatographic separation, together with a gradient elution of acetonitrile in 5 mM ammonium formate buffer (pH 3) as mobile phasw. After ionisation in the pneumatically-assisted electrospray source and in-soure collision induced dissociation, acquisition was performed in the selected ion monitoring mode. Recoveries were 69 and 47% on average, detection limits 2.5 and 0.25 ng/ml and quantitation limits 10 and 0.5 ng/ml for irinotecan and SN-38, respectively. Reproducibility was good and the method was linear from limits of quantitation up to 10 000 ng/ml for irinotecan, and up to 100 ng/ml for SN-38. This sensitive and highly specific method is suitable both for pharmacokinetic studies and routine therapeutic drug monitoring.
机译:提出了两种灵敏,准确的液相色谱-电喷雾质谱(LC-ES-MS)方法,使用相同的色谱和检测条件测定伊立替康及其活性代谢物SN-38的总含量。两者均以喜树碱为内标(I.S.)。伊立替康的样品预处理涉及简单的乙腈沉淀蛋白,而SN-38则需要液-液萃取。色谱分离使用Symmetry C_(18),3.5#μm(150×1 mm ID)反相柱进行色谱分离,并在5 mM甲酸铵缓冲液(pH 3)中将乙腈进行梯度洗脱,作为流动相。在气动辅助电喷雾源中进行电离并在源内碰撞诱导的解离后,以选定的离子监测模式进行采集。伊立替康和SN-38的平均回收率分别为69%和47%,检出限分别为2.5和0.25 ng / ml和定量限为10和0.5 ng / ml。重现性好,从伊利替康的定量限到10000 ng / ml,从SN-38的定量限到100 ng / ml,该方法是线性的。这种灵敏且高度特异性的方法适用于药代动力学研究和常规治疗药物监测。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号