首页> 外文期刊>Clinical Biochemistry >Verification of an immunoturbidimetric assay for heart-type fatty acid-binding protein (H-FABP) on a clinical chemistry platform and establishment of the upper reference limit
【24h】

Verification of an immunoturbidimetric assay for heart-type fatty acid-binding protein (H-FABP) on a clinical chemistry platform and establishment of the upper reference limit

机译:在临床化学平台上验证心脏型脂肪酸结合蛋白(H-FABP)的免疫比浊测定法和参考上限的建立

获取原文
获取原文并翻译 | 示例

摘要

Background: Heart-type fatty acid-binding protein (H-FABP) is an early biomarker of cardiac injury. Randox Laboratories developed an immunoturbidimetric H-FABP assay for non-proprietary automated clinical chemistry analysers that could be useful in the emergency department. We verified the analytical performances claimed by Randox Laboratories on Roche Cobas 6000 clinical chemistry platform in use in our laboratory, and we defined our own 99th percentile upper reference limit for H-FABP.
机译:背景:心脏型脂肪酸结合蛋白(H-FABP)是心脏损伤的早期生物标志物。 Randox实验室为非专有的自动化临床化学分析仪开发了一种免疫浊度H-FABP测定法,可用于急诊科。我们验证了Randox实验室在我们实验室中使用的Roche Cobas 6000临床化学平台上声称的分析性能,并定义了我们自己的H-FABP第99个百分位数上限。