An acid hydrolysis method for quantification of plasma free and total carnitine by flow injection tandem mass spectrometry.

机译:一种通过流动注射串联质谱法定量测定血浆中游离和总肉碱含量的酸水解方法。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVES: To develop a tandem mass spectrometry (MS/MS) method for plasma free and total carnitine, incorporating acid hydrolysis for total carnitine measurement, for improved accuracy and sample throughput. DESIGN AND METHODS: Limit of detection (LOD), limit of quantitation (LOQ), linearity, accuracy, precision coefficient of variation (CV), method comparison with a radioenzymatic assay (REA) and liquid chromatography tandem mass spectrometry assay (LC-MS/MS) were performed. The method was tested for interferences and for long-term performance. RESULTS: The method LOD and LOQ were 0.87 and 1.36 mumol/L for free carnitine and 1.79 and 2.54 mumol/L for total carnitine, respectively. No interferences were detected. Long-term measurement CVs for both free and total carnitine were <4%. CONCLUSIONS: The method shows acceptable performance characteristics with improved speed and low level accuracy compared to existing REA and MS/MS methods.
机译:目的:开发一种用于无血浆和总肉碱的串联质谱(MS / MS)方法,并结合酸水解法进行总肉碱的测量,以提高准确性和样品通量。设计与方法:检测限(LOD),定量限(LOQ),线性,准确度,精密变异系数(CV),与放射酶法(REA)和液相色谱串联质谱法(LC-MS)的方法比较/ MS)。测试了该方法的干扰和长期性能。结果:游离肉碱的LOD和LOQ分别为0.87和1.36 mumol / L,总肉碱的LOD和LOQ分别为1.79和2.54 mumol / L。未检测到干扰。游离肉碱和总肉碱的长期测量CV均小于4%。结论:与现有的REA和MS / MS方法相比,该方法显示出可接受的性能特征,具有改进的速度和较低的水平准确度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号