首页> 外文期刊>Clinical Biochemistry >Head-to-head comparison of an automated immunometric and an automated HPLC method for the quantification of HbA1c.
【24h】

Head-to-head comparison of an automated immunometric and an automated HPLC method for the quantification of HbA1c.

机译:自动定量和自动HPLC方法进行HbA1c定量的头对头比较。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVES: To compare a fully automated immunoassay and a fully automated HPLC method for the quantification of HbA1c. DESIGN AND METHODS: For an inter-method comparison, 200 consecutive clinical samples were analyzed. Imprecision was studied with pooled whole blood. RESULTS: Very close agreement of results (identical mean result (6.0%); r=0.99), and very similar reproducibility (total CV approximately 1%) were observed. CONCLUSIONS: Both assays offer very similar analytical performance, although the analytical techniques used are completely different.
机译:目的:比较全自动免疫测定法和全自动HPLC方法定量HbA1c的方法。设计与方法:为了进行方法间比较,分析了200个连续的临床样本。对不精确性进行了全血合并研究。结果:结果非常一致(平均结果相同(6.0%); r = 0.99),并且再现性非常相似(总CV约1%)。结论:尽管所使用的分析技术完全不同,但两种分析均提供非常相似的分析性能。