首页> 外文期刊>Journal of cardiovascular electrophysiology >T-wave oversensing in implantable cardiac defibrillators is due to technical failure of device sensing.
【24h】

T-wave oversensing in implantable cardiac defibrillators is due to technical failure of device sensing.

机译:植入式心脏除颤器中的T波感应过度是由于设备感应技术故障所致。

获取原文
获取原文并翻译 | 示例
           

摘要

The expanded indications for the use of implantable cardioverter defibrillators (ICDs) for primary prevention of sudden cardiac death have increased concerns about inappropriate delivery of therapies. Dual-chamber systems have improved the capacity of ICDs to differentiate ventricular arrhythmias from supraventricular arrhythmias. Still, T-wave oversensing is a major source of inappropriate therapy. It is likely the true incidence of T-wave oversensing is greater than reported as documented events reflect only those stored in a device memory. Reviewing cases of T-wave oversensing that failed noninvasive correction; we found successful resolution resulted from generator replacement. We conclude that T-wave oversensing is due to inadequate signal processing by some ICD generators.
机译:使用植入式心脏复律除颤器(ICD)来预防心源性猝死的广泛适应症增加了人们对不适当治疗的担忧。双腔系统提高了ICD区分室性心律失常和室上性心律失常的能力。尽管如此,T波过度感应仍是治疗不当的主要原因。由于记录的事件仅反映存储在设备存储器中的事件,因此T波超感的真实发生率可能大于报告的发生率。复查无创矫正失败的T波超视情况;我们发现发电机更换成功解决了问题。我们得出结论,T波超感是由于某些ICD发生器对信号的处理不足所致。