首页> 外文期刊>Journal of cardiovascular electrophysiology >Implantable Cardioverter-Defibrillator Malfunction due to Mechanical Failure of the Header Connection.
【24h】

Implantable Cardioverter-Defibrillator Malfunction due to Mechanical Failure of the Header Connection.

机译:割台连接器机械故障引起的植入式心脏复律除颤器故障。

获取原文
获取原文并翻译 | 示例

摘要

ICD Malfunction. Four cases of ICD system malfunction occurring within the Medtronic Marquis family over a limited time frame are reported. These cases shared a common element of defective connections between the header post and the right ventricular pace/sense terminal pin that required intervention and reconnection. Inappropriate shocks occurred in two cases. The circumstances of these cases provide important information about the potential difficulties with these devices and highlight the deficiencies of the current system in tracking and analyzing device-related problems. (J Cardiovasc Electrophysiol, Vol. 15, pp. 1-5, September 2004)
机译:ICD故障。据报道,在有限的时间内在Medtronic Marquis系列中发生了四起ICD系统故障的情况。这些情况共有一个共同的要素,就是集管柱和右心室起搏/感知端子引脚之间的连接不良,需要干预和重新连接。有两种情况发生不当的电击。这些情况的情况提供了有关这些设备潜在困难的重要信息,并突出了当前系统在跟踪和分析与设备有关的问题方面的缺陷。 (J Cardiovasc Electrophysiol,Vol.15,pp.1-5,2004年9月)