Biventricular shocking leads improve defibrillation efficacy.

机译:双室电击导线可改善除颤功效。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

INTRODUCTION: A single lead active can configuration has been widely used in patients with life-threatening ventricular arrhythmias. Occasionally, however, such a defibrillation lead configuration may not achieve adequate defibrillation threshold (DFT). The purpose of this study was to determine whether addition of a left ventricular (LV) lead can improve defibrillation efficacy. METHODS AND RESULTS: Three transvenous defibrillation leads (8.3-French with a 5-cm long unipolar coil) were placed in the right ventricle (RV), LV, and superior vena cava (SVC), along with an active can (92 cm2) in the left subpectoral area. The DFT stored energy of seven combinations of these defibrillation leads were compared in a pig ventricular fibrillation model using a biphasic defibrillation waveform (125 microF, 6.5/3.5 msec). A biventricular leads active can configuration in which the RV and LV leads were of the same polarity reduced the DFT stored energy by approximately 35% when compared to a single RV lead active can configuration (9.6 +/- 3.0 J vs 15.0 +/- 7.2 J, respectively, P = 0.02). Moreover, adding a SVC lead further reduced the DFT energy (8.4 +/- 3.3 J). CONCLUSION: A biventricular leads active can configuration can significantly improve defibrillation efficacy as compared to a single lead active can configuration. In such a defibrillation lead configuration, the polarity of RV and LV leads should be the same.
机译:简介:单个铅活性罐配置已被广泛用于危及生命的室性心律失常患者。但是,有时,这种除颤导线配置可能无法达到足够的除颤阈值(DFT)。这项研究的目的是确定是否添加左心室(LV)导线可以提高除颤功效。方法和结果:将三根静脉除颤导线(8.3法式,带5 cm长的单极线圈)放置在右心室(RV),LV和上腔静脉(SVC)中,并插入一个活动罐(92 cm2)。在左胸下区域。使用双相除颤波形(125 microF,6.5 / 3.5 msec)在猪心室纤颤模型中比较了这7种除颤导线组合的DFT储能。与单个RV引线有源罐配置相比,RV和LV引线具有相同极性的双心室引线有源罐配置将DFT存储的能量减少了约35%(9.6 +/- 3.0 J与15.0 +/- 7.2) J分别为P = 0.02)。此外,添加SVC引线进一步降低了DFT能量(8.4 +/- 3.3 J)。结论:双心室引线活性罐配置与单根引线活性罐配置相比可显着提高除颤功效。在这种除颤导线配置中,RV和LV导线的极性应相同。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号