Cardiovascular Perforation During Placement of an Avalon Elite (R) Bicaval Dual Lumen ECMO Cannula in a Newborn

机译:在新生儿中放置Avalon Elite(R)Bicaval双腔ECMO套管时的心血管穿孔

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Proper functioning of the Avalon Elite (R) bicaval dual lumen ECMO cannula (Maquet Cardiovascular, Wayne, NJ, USA) requires precise placement of the distal draining port within the lumen of the inferior vena cava (IVC). In order to advance the cannula to the correct position, a 0.038 guidewire is placed into the IVC under echocardiographic or fluoroscopic guidance. We report a case of perforation of the intrapericardiac section of the IVC by the guidewire in a neonate, not detected at the time of placement, resulting in cardiac tamponade and death. We recommend routine echocardiographic surveillance after placement of Avalon Elite (R) cannulas to rule out hemopericardium. doi: 10.1111/jocs.12507 (J Card Surg 2015;30:370-372)
机译:Avalon Elite(R)双腔双腔ECMO套管(Maquet Cardiovascular,韦恩,新泽​​西州,美国)的正常运行需要在下腔静脉(IVC)腔内精确放置远端引流口。为了将插管推进到正确的位置,将0.038导丝在超声心动图或荧光镜引导下放入IVC。我们报道了一例在新生儿中由导丝对IVC的心内膜穿孔的病例,在放置时未发现,导致心脏压塞和死亡。我们建议在放置Avalon Elite(R)插管后进行常规超声心动图监测,以排除心包积血。 doi:10.1111 / jocs.12507(J Card Surg 2015; 30:370-372)