Aim: To compare the international normalised ratio (INR) value of patients evaluated using the CoaguChek XS versus conventional laboratory methods, in the period after open-heart surgery for mechanical valve replacement until a therapeutic range is achieved using vitamin K antagonists (VKA) together with low molecular weight heparin (LMWH). Methods: One hundred and five patients undergoing open-heart surgery for mechanical valve replacement were enrolled. Blood samples were collected from patients before surgery, and on the second and fifth postoperative days, simultaneously for both the point of care device and conventional laboratory techniques. Patients were administered VKA together with LMWH at therapeutic doses (enoxaparin 100 IU/kg twice daily) subcutaneously, until an effective range was achieved on approximately the fifth day after surgery. Results: The mean INR values using the CoaguChek XS preoperatively and on the second and fifth days postoperatively were 1.20 (SD ± 0.09), 1.82 (SD ± 0.45), and 2.55 (SD ± 0.55), respectively. Corresponding results obtained using conventional laboratory techniques were 1.18 (SD ± 0.1), 1.81 (SD ± 0.43), and 2.51 (SD ± 0.58). The correlation coefficient was r = 0.77 preoperatively, r = 0.981 on postoperative day 2, and r = 0.983 on postoperative day 5. Discussion: Results using the CoaguChek XS Handheld Coagulation Analyzer correlated strongly with conventional laboratory methods, in the bridging period between open-heart surgery for mechanical valve replacement and the achievement of a therapeutic range on warfarin and LMWH.
机译:目的:为了比较使用CoaguChek XS和传统实验室方法评估的患者的国际标准化比率(INR),在进行心脏直视手术后进行机械瓣膜置换术之前,直到同时使用维生素K拮抗剂(VKA)达到治疗范围为止低分子量肝素(LMWH)。方法:纳入105例接受心脏直视手术以更换机械瓣膜的患者。在手术前,术后第二天和第五天同时从护理设备和常规实验室技术中采集患者的血样。皮下给予治疗剂量(依诺肝素100 IU / kg,每天两次)的VKA和LMWH,直到术后约第五天达到有效范围。结果:术前和术后第二天和第五天使用CoaguChek XS的平均INR值分别为1.20(SD±0.09),1.82(SD±0.45)和2.55(SD±0.55)。使用常规实验室技术获得的相应结果是1.18(SD±0.1),1.81(SD±0.43)和2.51(SD±0.58)。术前相关系数为r = 0.77,术后第2天为r = 0.981,术后第5天为r = 0.983。机械瓣膜置换术的心脏外科手术,并取得了华法林和LMWH的治疗范围。

著录项

相似文献

  • 外文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号