首页> 外文期刊>Journal of cardiac failure >Screening for Outflow Cannula Malfunction of Left Ventricular Assist Devices (LVADs) With the Use of Doppler Echocardiography: New LVAD-Specific Reference Values for Contemporary Devices
【24h】

Screening for Outflow Cannula Malfunction of Left Ventricular Assist Devices (LVADs) With the Use of Doppler Echocardiography: New LVAD-Specific Reference Values for Contemporary Devices

机译:多普勒超声心动图筛查左心室辅助设备(LVAD)的流出套管功能不良:当代设备新的LVAD特定参考值

获取原文
获取原文并翻译 | 示例
       

摘要

Background: Echocardiographic assessment of left ventricular assist devices (LVADs) is used as a screening tool to evaluate the integrity and mechanics of the pump and circuit. We aimed to 1) establish the normal range and upper reference limit of peak velocity of the outflow cannula for the modern era of LVADs and 2) assess the clinical performance of the currently cited and newly proposed reference limits in patients with continuous-flow LVADs as a screening tool for cannula malfunction.
机译:背景:左心室辅助设备(LVAD)的超声心动图评估被用作筛选工具,以评估泵和回路的完整性和力学性能。我们的目标是:1)建立现代LVAD的流出套管峰值速度的正常范围和最高参考极限,以及2)评估当前引用和新提议的参考值在连续流LVAD患者中的临床表现,如下所示:套管故障的筛查工具。