Dentin permeability was measured alternatively with two methods: a 10-mu l capillary method with visual evaluation (PC) and a motorized automatic measuring device (Flodec, ED), both interposed in a simulated perfusion system. Eight human third molar coronal fragments were connected to systems, and their permeability to distilled water measured at 0, 5, 10, 15, 20, 25, and 29 cm H_2O pressure. Resultant permeabilities (in mu l/s) for both techniques were interrelated with the use of the Passing and Bablok nonparametric method, which gives information about the range of constant and proportional errors and their 95 percent confidence intervals (95CI). The relationship between the methods is described by the regression formula: FD = -0.0003 + 0.945 centre dot PC, with 95CI for constant (-0.0015-0.0009) and for slope (0.738-1.168), indicating that both methods are interchangeable, although not identical.
机译:交替使用两种方法测量牙本质通透性:具有可视评估的10微升毛细管法(PC)和电动自动测量装置(Flodec,ED),两者均插入模拟灌注系统中。将八个人类第三磨牙冠状碎片连接到系统,并在0、5、10、15、20、25和29 cm H_2O压力下测量其对蒸馏水的渗透性。两种技术的最终渗透率(以μl/ s为单位)与Passing和Bablok非参数方法的使用相关,该方法可提供有关恒定误差和比例误差的范围及其95%置信区间(95CI)的信息。方法之间的关系由回归公式描述:FD = -0.0003 + 0.945中心点PC,常数(-0.0015-0.0009)和斜率(0.738-1.168)的95CI,表明这两种方法可以互换,尽管不能互换相同。