Comparison of V-Loc? 180 wound closure device and Quill? PDO knotless tissue-closure device for intradermal closure in a porcine in vivo model: evaluation of biomechanical wound strength.

机译:比较V-Loc? 180伤口闭合装置和鹅毛笔? PDO无结组织闭合装置用于猪体内模型的皮内闭合:生物力学伤口强度评估。

获取原文
获取原文并翻译 | 示例
       

摘要

The objective of this study was to compare the biomechanical strength of two barbed suture devices: V-Loc? 180 Wound Closure Device and Quill? PDO Knotless Tissue-Closure Device following primary cosmetic skin closures in a porcine dermal model.This prospective randomized, controlled in vivo trial compared size 3/0 V-Loc? 180 device to size 2/0 Quill? PDO device. Both products were tested for dermal closure in adult porcine models and evaluated at five timepoints. At postoperative days 0, 3, 7, 14, and 28 sutured tissue regions were excised post mortem and tested for intradermal wound holding strength.Wounds closed with V-Loc? 180 device were stronger than Quill? PDO device at days 0, 3, 7, and 14 with these differences being significant (p < 0.05) at days 3 and 7. At day 3, the average maximum load of V-Loc? 180 was 13.53 kgf and Quill? PDO was 10.38 kgf (p = 0.002). At day 7, the average maximum load of V-Loc? 180 was 10.4 kgf and Quill? PDO was 7.56 kgf (p = 0.001). Throughout the duration of the study, there was no suture extrusion or tissue distortion and all wounds healed with no major complications.In this study, V-Loc? 180 device was significantly stronger than Quill? PDO device during the critical phases of wound healing in skin.
机译:这项研究的目的是比较两种带刺的缝合设备:V-Loc?的生物力学强度。 180伤口闭合装置和鹅毛笔? PDO无结组织闭合装置在猪皮肤模型中进行初级美容皮肤闭合后。该前瞻性随机对照体内试验比较了3/0 V-Loc大小。 180尺寸为2/0的设备吗? PDO设备。两种产品均在成年猪模型中进行了皮肤封闭测试,并在五个时间点进行了评估。在术后第0、3、7、14和28天,从死后切开缝合的组织区域,并测试其皮内伤口的保持力。用V-Loc封闭伤口。 180装置比羽毛笔强?在第0、3、7和14天的PDO设备,这些差异在第3天和第7天显着(p <0.05)。在第3天,V-Loc?的平均最大负载? 180是13.53 kgf和Quill? PDO为10.38 kgf(p = 0.002)。在第7天,V-Loc的平均最大负载? 180是10.4千克力和羽毛笔? PDO为7.56 kgf(p = 0.001)。在整个研究过程中,没有缝合线挤压或组织变形,所有伤口均愈合,没有重大并发症。 180装置明显比Quill强?在皮肤伤口愈合的关键阶段使用PDO设备。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号