首页> 外文期刊>Journal of biomechanical engineering. >In Vitro Pulsatility Analysis of Axial-Flow and Centrifugal-Flow Left Ventricular Assist Devices
【24h】

In Vitro Pulsatility Analysis of Axial-Flow and Centrifugal-Flow Left Ventricular Assist Devices

机译:轴流和离心流左心室辅助装置的体外搏动分析

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Recently, continuous-flow ventricular assist devices (CF-VADs) have supplanted older, pulsatile-flow pumps, for treating patients with advanced heart failure. Despite the excellent results of the newer generation devices, the effects of long-term loss of pulsatility remain unknown. The aim of this study is to compare the ability of both axial and centrifugal continuous-flow pumps to intrinsically modify pulsatility when placed under physiologically diverse conditions. Four VADs, two axial- and two centrifugal-flow, were evaluated on a mock circulatory flow system. Each VAD was operated at a constant impeller speed over three hypothetical cardiac conditions: normo-tensive, hypertensive, and hypotensive. Pulsatility index (PI) was compared for each device under each condition. Centrifugal-flow devices had a higher PI than that of axial-flow pumps. Under normo-tension, flow PI was 0.98 ± 0.03 and 1.50 ± 0.02 for the axial and centrifugal groups, respectively (p < 0.01). Under hypertension, flow PI was 1.90 ± 0.16 and 4.21 ± 0.29 for the axial and centrifugal pumps, respectively (p = 0.01). Under hypotension, PI was 0.73 ± 0.02 and 0.78 ± 0.02 for the axial and centrifugal groups, respectively (p = 0.13). All tested CF-VADs were capable of maintaining some pulsatile-flow when connected in parallel with our mock ventricle. We conclude that centrifugal-flow devices outperform the axial pumps from the basis of PI under tested conditions.
机译:最近,连续流心室辅助设备(CF-VAD)取代了较旧的脉动流泵,用于治疗晚期心力衰竭患者。尽管新一代设备具有出色的结果,但长期丧失搏动性的影响仍然未知。这项研究的目的是比较当放置在生理上不同的条件下时,轴向和离心式连续流泵在本质上改变脉动性的能力。在模拟循环流系统上评估了四个VAD,两个轴向流和两个离心流。在三种假设的心脏病条件下,每个VAD均以恒定的叶轮速度运行:血压正常,高血压和血压低。比较了每种设备在每种情况下的脉动指数(PI)。离心流装置的PI高于轴流泵的PI。在常压下,轴向和离心组的流动PI分别为0.98±0.03和1.50±0.02(p <0.01)。在高血压情况下,轴向泵和离心泵的流量PI分别为1.90±0.16和4.21±0.29(p = 0.01)。在低血压情况下,轴向和离心组的PI分别为0.73±0.02和0.78±0.02(p = 0.13)。当与我们的模拟心室并联时,所有测试的CF-VAD都能保持一定的脉动血流。我们得出的结论是,在测试条件下,基于PI的离心流量装置的性能优于轴向泵。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号